In 2020, the global prevalence of glaucoma was estimated to be 76 million and it was projected to increase to 111.8 million by 2040. Accurate intraocular pressure (IOP) measurement is imperative in glaucoma management since it is the only modifiable risk factor. Numerous studies have compared the reliability of IOP measured using transpalpebral tonometers and Goldmann applanation tonometry (GAT). This systematic review and meta-analysis aims to update the existing literature with a reliability and agreement comparison of transpalpebral tonometers against the gold standard GAT for IOP measurement among individuals presenting for ophthalmic examinations. The data collection will be performed using a predefined search strategy through electronic databases. Prospective methods-comparison studies published between January 2000 and September 2022 will be included. Studies will be deemed eligible if they report empirical findings on the agreement between transpalpebral tonometry and Goldmann applanation tonometry. The standard deviation and limits of agreement between each study and their pooled estimate along with weights and percentage of error will be reported using a forest plot. Cochrane's Q test and the I statistic will be used to assess heterogeneity, and the publication bias will be investigated using a funnel plot, Begg's and Egger's tests. The review results will provide additional evidence on the reliability of transpalpebral tonometers that, in turn, could possibly assist practitioners to make informed decision about using it as a screening or diagnostic device for clinical practice, outreach camps, or home-based screening. Institutional Ethics Committee registration number: RET202200390. PROSPERO Registration Number: CRD42022321693.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10391437PMC
http://dx.doi.org/10.4103/IJO.IJO_3321_22DOI Listing

Publication Analysis

Top Keywords

transpalpebral tonometers
16
goldmann applanation
12
reliability transpalpebral
8
systematic review
8
review meta-analysis
8
iop measurement
8
applanation tonometry
8
registration number
8
will
7
transpalpebral
5

Similar Publications

Topic: The systematic review and meta-analysis consolidates the extant body of evidence comparing reliability and agreement between Transpalpebral Tonometers (TT) and Goldmann Applanation Tonometer (GAT).

Clinical Relevance: With a global prevalence of 3.54 percent, glaucoma stands as the second leading cause of preventable blindness.

View Article and Find Full Text PDF

Purpose: We developed a novel transpalpebral self-tonometer called the TapEye tonometer (TET) based on palpation of the upper eyelid. Our goal was to evaluate a method for improving the accuracy of measuring intraocular pressure (IOP) through the eyelid.

Methods: Participants underwent standardized training by technicians and were required to be able to use the TET for study inclusion.

View Article and Find Full Text PDF
Article Synopsis
  • Eyelid spasms in patients with hemifacial spasm (HFS) may lead to elevated intraocular pressure (IOP), which complicates accurate measurement using traditional methods like the Goldmann applanation tonometer (GAT).
  • This study evaluated IOP using the Diaton® transpalpebral tonometer in 27 HFS patients before and after botulinum toxin type A (BTX-A) treatment and compared these results to GAT measurements.
  • The results indicated a significant reduction in IOP on the affected side after BTX-A treatment, showing that eyelid spasms can increase IOP, but no significant differences were found between Diaton® and GAT readings post-treatment.
View Article and Find Full Text PDF

Objectives: The objective of the study is to evaluate the agreement between Goldmann applanation tonometer (GAT) and Easyton transpalpebral tonometer, Tonopen, and Icare in patients with Keratoconus.

Methods: This cross-sectional study included 46 eyes of 26 patients with keratoconus. Intraocular pressure (IOP) is measured using easyton, icare, tonopen, and GAT.

View Article and Find Full Text PDF

Background: The mechanism that could increase intraocular pressure (IOP) during scleral lens (SL) wear is not fully understood, although it may be related to compression of the landing zone on structures involved in aqueous humor drainage.

Methods: Thirty healthy subjects were fitted with two SLs of different sizes (L1 = 15.8 mm, L2 = 16.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!