Introduction: We aimed to assess the validity and reproducibility of a wearable hydration device in a cohort of maintenance dialysis patients.
Methods: We conducted a prospective, single-arm observational study on 20 haemodialysis patients between January and June 2021 in a single centre. A prototype wearable infrared spectroscopy device, termed the Sixty device, was worn on the forearm during dialysis sessions and nocturnally. Bioimpedance measurements were performed 4 times using the body composition monitor (BCM) over 3 weeks. Measurements from the Sixty device were compared with the BCM overhydration index (litres) pre- and post-dialysis and with standard haemodialysis parameters.
Results: 12 out of 20 patients had useable data. Mean age was 52 ± 12.4 years. The overall accuracy for predicting pre-dialysis categories of fluid status using Sixty device was 0.55 [K = 0.00; 95% CI: -0.39-0.42]. The accuracy for the prediction of post-dialysis categories of volume status was low [accuracy = 0.34, K = 0.08; 95% CI: -0.13-0.3]. Sixty outputs at the start and end of dialysis were weakly correlated with pre- and post-dialysis weights ( = 0.27 and = 0.27, respectively), as well as weight loss during dialysis ( = 0.31), but not ultrafiltration volume ( = 0.12). There was no difference between the change in Sixty readings overnight and the change in Sixty readings during dialysis (mean difference 0.09 ± 1.5 kg), [(39) = 0.38, = 0.71].
Conclusion: A prototype wearable infrared spectroscopy device was unable to accurately assess changes in fluid status during or between dialysis sessions. In the future, hardware development and advances in photonics may enable the tracking of interdialytic fluid status.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10184568 | PMC |
http://dx.doi.org/10.1159/000529899 | DOI Listing |
Sci Rep
December 2024
Department of Ophthalmology, University of Health Sciences, Antalya Education and Research Hospital, Antalya, 07050, Turkey.
Our current prospective cross-sectional study aimed to investigate the effect of anti-vascular endothelial growth factor (VEGF) drugs used in the treatment of retinopathy of prematurity on retinal maturation and persistent avascular retina (PAR). Retinal imaging was performed with Optos confocal laser ophthalmoscopy for 100 patients aged 4 to 8 years who were screened and treated for retinopathy of prematurity (ROP) during the neonatal period. The ROP examination findings (stage and zone) and treatment history (age in weeks at time of treatment and anti-VEGF drug used) from the neonatal period were reviewed.
View Article and Find Full Text PDFBMC Oral Health
December 2024
Faculty of Odonto-Stomatology, University of Health Sciences, Vietnam National University Ho Chi Minh City, Ho Chi Minh City, Vietnam.
Background: The success of a restoration largely depends on the quality of its fit. This study aimed to investigate the fit quality of monolithic zirconia veneers (MZVs) produced through traditional and digital workflows.
Methods: A typodont maxillary right central incisor was prepared.
Lasers Med Sci
December 2024
Department of Urology Surgery, Beijing Traditional Chinese Medicine Hospital Affiliated to Capital Medical University, No.23 Art Museum Back Street, Dongcheng District, Beijing, 100010, China.
To compare the efficacy and safety of low-power holmium laser enucleation of the prostate (LP-HoLEP) with plasma kinetic resection of prostate (PKRP). Sixty-three patients treated with transurethral LP-HoLEP (observation group) and 68 patients treated with transurethral PKRP (control group) at Beijing Hospital of Traditional Chinese Medicine from November 2019 to November 2022 were retrospectively compared with regard to operation duration, intra-operative blood loss, prostate resection ratio, postoperative bladder irrigation time, postoperative indwelling urinary catheter time, postoperative urinary incontinence incidence, International Prostate Symptom Scale (IPSS), maximum urine flow rate (Qmax), and residual urine volume (RUV). In both groups, postoperative IPSS, Qmax, and RUV were significantly improved compared to preoperative values (P < 0.
View Article and Find Full Text PDFArch Orthop Trauma Surg
December 2024
Department of Orthopaedic Surgery, Sint Maartenskliniek, Ubbergen, The Netherlands.
Introduction: Recent studies have increasingly provided evidence that one-stage septic revisions for hip and knee are a safe treatment option for periprosthetic joint infections (PJI) in selected patients. However, there is still a wide treatment variation concerning indications and execution among different practices. This study aimed to describe these differences.
View Article and Find Full Text PDFBMC Surg
December 2024
Department of Ultrasound, The Seventh Affiliated Hospital of Sun Yat-Sen University, #628 Zhenyuan Road, Shenzhen, Guangdong, 518100, China.
Purpose: The aim of this study was to determine whether ultrasound guidance can reduce the duration, blood loss volume and invasiveness of surgery for internal fixator removal.
Methods: The clinical data from 35 adults patients who underwent ultrasound-guided surgical removal of internal fixators after complete healing of limb fractures between June 2019 and April 2023 were retrospectively analysed and compared with those from 34 controls who underwent the procedure without ultrasound guidance. Data concerning the patients' demographic and clinical characteristics and surgical sites were collected.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!