A Novel Ultrahigh-Dose-Rate Proton Therapy Technology: Spot-Scanning Proton Arc Therapy + FLASH (SPLASH).

Int J Radiat Oncol Biol Phys

Department of Radiation Oncology, Beaumont Health System, Royal Oak, Michigan. Electronic address:

Published: November 2023

Purpose: To take full advantage of FLASH dose rate (40 Gy/s) and high-dose conformity, we introduce a novel optimization and delivery technique, the spot-scanning proton arc therapy (SPArc) + FLASH (SPLASH).

Methods And Materials: SPLASH framework was implemented in an open-source proton planning platform (MatRad, Department of Medical Physics in Radiation Oncology, German Cancer Research Center). It optimizes with the clinical dose-volume constraint based on dose distribution and the dose-average dose rate by minimizing the monitor unit constraint on spot weight and accelerator beam current sequentially, enabling the first dynamic arc therapy with voxel-based FLASH dose rate. This new optimization framework minimizes the overall cost function value combined with plan quality and voxel-based dose-rate constraints. Three representative cases (brain, liver, and prostate cancer) were used for testing purposes. Dose-volume histogram, dose-rate-volume histogram, and dose-rate map were compared among intensity modulated proton radiation therapy (IMPT), SPArc, and SPLASH.

Results: SPLASH/SPArc could offer superior plan quality over IMPT in terms of dose conformity. The dose-rate-volume histogram results indicated SPLASH could significantly improve V Gy/s in the target and region of interest for all tested cases compared with SPArc and IMPT. The optimal beam current per spot is simultaneously generated, which is within the existing proton machine specifications in the research version (<200 nA).

Conclusions: SPLASH offers the first voxel-based ultradose-rate and high-dose conformity treatment using proton beam therapy. Such a technique has the potential to fit the needs of a broad range of disease sites and simplify clinical workflow without applying a patient-specific ridge filter, which has never before been demonstrated.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ijrobp.2023.05.012DOI Listing

Publication Analysis

Top Keywords

dose rate
12
spot-scanning proton
8
proton arc
8
flash dose
8
arc therapy
8
beam current
8
plan quality
8
dose-rate-volume histogram
8
proton
6
dose
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!