AI Article Synopsis

  • Research shows that short dental implants with platform matching are useful for restoring atrophic jaws when standard implants can't be used without bone work.
  • This study evaluates the risk of technical failures using finite element analysis for an all-on-4 configuration with different types of short dental implants in patients with atrophic mandibles.
  • Results indicate that the straight standard implant configuration had the lowest stress values, while the tilted and shorter implants showed higher stress levels on various components and the surrounding bone.

Article Abstract

Short dental implants with platform matching connection have been used for the rehabilitation of atrophic jaws whenever standard-length dental implants cannot be placed without prior bone augmentation. Yet, there remains a lack of data regarding the risk of technical failures when the all-on-4 configuration is performed in atrophic jaws with platform-switching distal short dental implants. Thus, the current study used the finite element method to evaluate the mechanical behavior at the level of the prosthetic components of the all-on-4 concept performed in atrophic mandible using short-length distal implants with platform switching (PSW) connection. Three models of the all-on-4 configuration were generated in human atrophic mandibles. The geometric models consisted of PSW connection tilted standard (AO4T; θ = 30 deg; 11 mm-length), straight standard (AO4S; θ = 0 deg; 11 mm-length) and straight short (AO4Sh; θ = 0 deg; 8 mm-length) distal implants. A resultant force of 300 N was performed obliquely in the left side and posterior region of the prosthetic bar. The von Mises equivalent stress (σvm) and maximum and minimum principal stresses (σmax and σmin) were performed at level of the prosthetic components/implants and peri-implant bone crest, respectively. The general displacement of the models was also evaluated. The stress analysis was performed on the side of load application. The AO4S configuration showed the lowest values of σvm in the mesial left (ML) and distal left (DL) abutments (37.53 MPa and 232.77 MPa, respectively) and dental implants (91.53 MPa and 231.21 MPa, respectively). The AO4Sh configuration showed the highest values of σvm in the bar screw (102.36 MPa), abutment (117.56 MPa), and dental implant (293.73 MPa) of the ML area. Among the models, the highest values of σmax and σmin were noticed in the peri-implant bone crest of the AO4T design (131.48 MPa and 195.31 MPa, respectively). All models showed similar values of general displacements, which were concentrated in the mandible symphysis. The all-on-4 configurations designed with PSW connection and tilted standard (AO4T; θ = 30 deg; 11 mm-length), straight standard (AO4S; θ = 0 deg; 11 mm-length) or straight short (AO4Sh; θ = 0 deg; 8 mm-length) distal implants were not associated with higher odds of technical failures. The AO4Sh design may be a promising option for the prosthetic rehabilitation of atrophic jaws.

Download full-text PDF

Source
http://dx.doi.org/10.1115/1.4062540DOI Listing

Publication Analysis

Top Keywords

dental implants
20
11 mm-length straight
16
short dental
12
atrophic jaws
12
distal implants
12
psw connection
12
platform-switching distal
8
distal short
8
implants
8
implants platform
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!