Purpose: The Medical Physics Working Group of the Radiation Therapy Study Group at the Japan Clinical Oncology Group is currently developing a virtual audit system for intensity-modulated radiation therapy dosimetry credentialing. The target dosimeters include films and array detectors, such as ArcCHECK (Sun Nuclear Corporation, Melbourne, Florida, USA) and Delta4 (ScandiDos, Uppsala, Sweden). This pilot study investigated the feasibility of our virtual audit system using previously acquired data.

Methods: We analyzed 46 films (32 and 14 in the axial and coronal planes, respectively) from 29 institutions. Global gamma analysis between measured and planned dose distributions used the following settings: 3%/3 mm criteria (the dose denominator was 2 Gy), 30% threshold dose, no scaling of the datasets, and 90% tolerance level. In addition, 21 datasets from nine institutions were obtained for array evaluation. Five institutions used ArcCHECK, while the others used Delta4. Global gamma analysis was performed with 3%/2 mm criteria (the dose denominator was the maximum calculated dose), 10% threshold dose, and 95% tolerance level. The film calibration and gamma analysis were conducted with in-house software developed using Python (version 3.9.2).

Results: The means ± standard deviations of the gamma passing rates were 99.4 ± 1.5% (range, 92.8%-100%) and 99.2 ± 1.0% (range, 97.0%-100%) in the film and array evaluations, respectively.

Conclusion: This pilot study demonstrated the feasibility of virtual audits. The proposed virtual audit system will contribute to more efficient, cheaper, and more rapid trial credentialing than on-site and postal audits; however, the limitations should be considered when operating our virtual audit system.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10243313PMC
http://dx.doi.org/10.1002/acm2.14040DOI Listing

Publication Analysis

Top Keywords

virtual audit
20
audit system
20
radiation therapy
12
pilot study
12
gamma analysis
12
system intensity-modulated
8
intensity-modulated radiation
8
japan clinical
8
clinical oncology
8
oncology group
8

Similar Publications

Objective: Babies born between 27 and 31 weeks of gestation contribute substantially towards infant mortality and morbidity. In England, their care is delivered in maternity services colocated with highly specialised neonatal intensive care units (NICU) or less specialised local neonatal units (LNU). We investigated whether birth setting offered survival and/or morbidity advantages to inform National Health Service delivery.

View Article and Find Full Text PDF

Context: A patient-centered care (PCC) environment allows athletic trainers (ATs) to develop trusting relationships with patients, enabling them to make the most informed care decisions. To provide PCC, the AT should assess health literacy and deliver quality patient education.

Objective: To explore the lived experiences of ATs from different job settings to identify how they deliver PCC specific to health literacy and patient education.

View Article and Find Full Text PDF

Background: The deep inferior epigastric perforator (DIEP) flap provides an effective and popular means for autologous breast reconstruction. However, with the complexity of the pathway, the environmental impact of the pathway has yet to be evaluated.

Methods: A retrospective analysis of 42 unilateral DIEPs at a single reconstructive center was performed.

View Article and Find Full Text PDF

Background: Clinician electronic actions within the electronic health record (EHR) are captured seamlessly in real-time during regular work activities in all major EHRs. Analysis of this EHR use metadata, such as audit log data, is increasingly used to understand the impact of work design on critical patient, workforce, and organizational outcomes.

Objective: Understand experiences and perspectives influencing the use and implementation of audit log data into practice.

View Article and Find Full Text PDF

Introduction: Trauma and orthopedics departments have traditionally used face-to-face (FTF) fracture clinics for non-operative fractures. Developed in 2011, the virtual fracture clinic (VFC) was fully implemented at an institution during the COVID-19 pandemic to reduce in-person interactions.

Aims: First, the study aims to measure the percentage of non-operative patients triaged through the VFC when this was optional and re-audit after implementing a COVID-19-related policy change mandating VFC triage.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!