Among the blood counts, platelet count is most often reported with inconsistency. Many of the analyzers work on electrical impedance principle for red blood cell (RBC) and platelet counting. However, with this technology, factors such as fragmented RBCs, microcytes, cytoplasmic fragments of leukemic cells, lipid particles, fungal yeast forms, and bacteria are known to interfere with platelet count and give spuriously elevated platelet counts. A 72-year-old male was admitted for the treatment of dengue infection who had serial platelet count monitoring. He had an initial platelet count of 48,000/cumm which suddenly improved to 2.6 lakhs within 6 h without any platelet transfusion. Peripheral smear however did not correlate with the machine-derived count. Repeat test after 6 h yielded a result of 56,000/cumm which correlated well with the peripheral smear. This falsely elevated count was due to the presence of lipid particles as the sample was drawn in the postprandial state.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10180802PMC
http://dx.doi.org/10.4103/ajts.ajts_65_21DOI Listing

Publication Analysis

Top Keywords

platelet count
20
platelet
8
lipid particles
8
peripheral smear
8
count
6
count impedance-based
4
impedance-based hematology
4
hematology analyzer
4
analyzer beware
4
beware trap!
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!