Characterization of 250 MeV Protons from the Varian ProBeam PBS System for FLASH Radiation Therapy.

Int J Part Ther

Department of Radiation Oncology and Winship Cancer Institute, Emory University, Atlanta, GA, USA.

Published: March 2023

Shoot-through proton FLASH radiation therapy has been proposed where the highest energy is extracted from a cyclotron to maximize the dose rate (DR). Although our proton pencil beam scanning system can deliver 250 MeV (the highest energy), this energy is not used clinically, and as such, 250 MeV has yet to be characterized during clinical commissioning. We aim to characterize the 250-MeV proton beam from the Varian ProBeam system for FLASH and assess the usability of the clinical monitoring ionization chamber (MIC) for FLASH use. We measured the following data for beam commissioning: integral depth dose curve, spot sigma, and absolute dose. To evaluate the MIC, we measured output as a function of beam current. To characterize a 250 MeV FLASH beam, we measured (1) the central axis DR as a function of current and spot spacing and arrangement, (2) for a fixed spot spacing, the maximum field size that achieves FLASH DR (ie, > 40 Gy/s), and (3) DR reproducibility. All FLASH DR measurements were performed using an ion chamber for the absolute dose, and irradiation times were obtained from log files. We verified dose measurements using EBT-XD films and irradiation times using a fast, pixelated spectral detector. R90 and R80 from integral depth dose were 37.58 and 37.69 cm, and spot sigma at the isocenter were σ = 3.336 and σ = 3.332 mm, respectively. The absolute dose output was measured as 0.343 Gy*mm/MU for the commissioning conditions. Output was stable for beam currents up to 15 nA and gradually increased to 12-fold for 115 nA. Dose and DR depended on beam current, spot spacing, and arrangement and could be reproduced with 6.4% and 4.2% variations, respectively. Although FLASH was achieved and the largest field size that delivers FLASH DR was determined as 35 × 35 mm, the current MIC has DR dependence, and users should measure dose and DR independently each time for their FLASH applications.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10166018PMC
http://dx.doi.org/10.14338/IJPT-22-00027.1DOI Listing

Publication Analysis

Top Keywords

250 mev
16
absolute dose
12
spot spacing
12
flash
10
dose
9
varian probeam
8
system flash
8
flash radiation
8
radiation therapy
8
highest energy
8

Similar Publications

Radiation Damage Mitigation in FeCrAl Alloy at Sub-Recrystallization Temperatures.

Materials (Basel)

December 2024

Department of Mechanical Engineering, The Pennsylvania State University, University Park, PA 16803, USA.

Traditional defect recovery methods rely on high-temperature annealing, often exceeding 750 °C for FeCrAl. In this study, we introduce electron wind force (EWF)-assisted annealing as an alternative approach to mitigate irradiation-induced defects at significantly lower temperatures. FeCrAl samples irradiated with 5 MeV Zr ions at a dose of 10 cm were annealed using EWF at 250 °C for 60 s.

View Article and Find Full Text PDF

Runaway electrons, accelerated in a tokamak discharge to high energies (tens of MeV), can cause serious damage to plasma facing components. Therefore, it is important to develop effective mitigation strategies to reduce the risk of tokamak damage. To study the effects of various mitigation strategies, a dedicated diagnostic, the calorimetry probe, was developed at the COMPASS tokamak.

View Article and Find Full Text PDF

Background: Pre-clinical studies demonstrate that delivering a high dose at a high dose rate result in less toxicity while maintaining tumor control, known as the FLASH effect. In proton therapy, clinical trials have started using 250 MeV transmission beams and more trials are foreseen. A novel aspect of FLASH treatments, compared to conventional radiotherapy, is the importance of dose rate next to dose and geometry.

View Article and Find Full Text PDF

Probing -wave superconductivity in UTe via point-contact junctions.

NPJ Quantum Mater

November 2024

Department of Physics, Maryland Quantum Materials Center, University of Maryland, College Park, MD USA.

Article Synopsis
  • Uranium ditelluride (UTe) is considered a top candidate for a -wave superconductor in bulk form, prompting detailed spectroscopic research.
  • Conductance measurements were taken through point-contact junctions at low temperatures (down to 250 mK) and high magnetic fields (up to 18 T), utilizing the Blonder-Tinkham-Klapwijk model for analysis.
  • The findings indicate a dominant -wave gap function with an amplitude of 0.26 ± 0.06 meV, supporting the idea of spin-triplet pairing in UTe's superconducting state.
View Article and Find Full Text PDF

In the present work, a comparative study on eco-friendly synthesis of zinc oxide (ZnO) sample 1 and sample 2 with 3.17 and 4.17 M NaOH, respectively, is reported.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!