BACKGROUND : Outbreaks of multidrug-resistant bacteria due to contaminated duodenoscopes and infection risks during the COVID-19 pandemic have driven the development of single-use endoscopes. The first single-use gastroscope is now available in Europe. Besides waste disposal and cost issues, the infection risk and performance remain unclear. We aimed to evaluate a single-use gastroscope in patients with signs of upper gastrointestinal bleeding. METHODS : 20 consecutive patients presenting with clinical signs of upper gastrointestinal bleeding between October and November 2022 were included in this case series. The primary aim was technical success, defined as access to the descending duodenum and adequate assessment of the upper gastrointestinal tract for the presence of a bleeding site. RESULTS : The primary aim was achieved in 19/20 patients (95 %). The bleeding site was identified in 18 patients. A therapeutic intervention was performed in six patients (two cap-mounted clips, one standard hemostatic clip, two variceal band ligations, one hemostatic powder, two adrenaline injections); technical and clinical success were achieved in all six patients. Two crossovers to a standard gastroscope occurred. CONCLUSIONS : Use of single-use gastroscopes may be feasible for patients presenting for urgent endoscopic evaluation and treatment of upper gastrointestinal bleeding.