Reducing the time delay of oxygen transport to the neonate on continuous positive airway pressure support: A bench study.

Front Pediatr

Department of Biomedical Technology, Faculty of Biomedical Engineering, Czech Technical University in Prague, Kladno, Czech Republic.

Published: April 2023

Background: Premature newborns often require oxygen support as part of their therapy. Systems for oxygen administration are developed to assure adequate oxygenation of newborns. Several factors were identified in the systems that contribute to the time delay between the change in the set inspiratory oxygen fraction and its actual delivery to tissues. In this study, we aimed to reduce the physical delay in oxygen delivery to newborns.

Methods: We developed an O Flush System (O-FS) that brings the source of oxygen as close to a patient as possible to make oxygen available for rapid delivery that compensates for the physical delay in the ventilator circuit. The O-FS system is built around an electromechanical on/off valve. We validated the O-FS concept in experiments with non-invasive Continuous Positive Airways Pressure (CPAP) ventilators.

Results: The O-FS accelerated oxygen delivery with all the tested systems and arrangements, typically by 5-15 s. We also observed that the application of supplemental oxygen increased the pressure in the ventilator circuit by 3-4 cmHO which may mitigate the apneic pauses that are common in premature newborns.

Conclusions: The O-FS system may work as a universal accessory of the CPAP lung ventilator and shorten the distribution of oxygen to the patient during oxygen desaturation events, possibly eliminating or interrupting apneic pauses in neonates, for whom oxygen therapy is an essential treatment. In clinical practice, the O-FS could help maintain normoxemic saturation values through adequate oxygen dosing in preterm neonates, thus reducing morbidity and mortality.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10155780PMC
http://dx.doi.org/10.3389/fped.2023.1141432DOI Listing

Publication Analysis

Top Keywords

oxygen
13
time delay
8
delay oxygen
8
continuous positive
8
physical delay
8
oxygen delivery
8
patient oxygen
8
ventilator circuit
8
o-fs system
8
apneic pauses
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!