Background: Diagnosis of Human T-cell Lymphotropic Virus (HTLV) types I and II infection requires sequencial testing with firstly a screening using an Enzyme immunoassay followed by a confirmatory test.
Objectives: To compare the performances of the Alinity i rHTLV-I/II (Abbott®) and LIAISON® XL murex recHTLV-I/II serological screening tests to the ARCHITECT rHTLVI/II test followed if positive by HTLV BLOT 2.4, MP Diagnostics as the reference.
Study Design: 119 serum samples from 92 known HTLV-I infected patients and 184 from uninfected patients with HTLV were analyzed in parallel with, Alinity i rHTLV-I/II, LIAISON® XL murex recHTLV-I/II and ARCHITECT rHTLVI/II.
Results: Alinity i rHTLV-I/II and LIAISON® XL murex recHTLV-I/II exhibited a total agreement with ARCHITECT rHTLVI/II for both positive and negative samples. Both tests are suitable alternatives for HTLV screening.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jcv.2023.105446 | DOI Listing |
Vox Sang
March 2024
Laboratório de Sorologia, Fundação HEMOMINAS, Belo Horizonte, Brazil.
Background And Objectives: Serological HTLV-1/2 screening is mandatory for blood donor candidates in Brazil. Our objective was to analyse HTLV test results in blood donors submitted for screening and confirmatory assays in a Brazilian blood bank.
Materials And Methods: Retrospective analysis (2017-2022) results of chemiluminescence immunoassays and confirmatory tests for HTLV-1/2 in reactive donors were performed.
J Clin Virol
July 2023
AP-HP, Hôpitaux Universitaires La Pitié Salpêtrière-Charles Foix, Service de Virologie, F-75013 Paris, France; Université Paris cité, INSERM UMR-S 1139 Physiopathologie et pharmacotoxicologie placentaire humaine, microbiote pré & post-natal, F-75006 Paris, France. Electronic address:
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!