Background: Interest in spatial fractionation radiotherapy has exponentially increased over the last decade as a significant reduction of healthy tissue toxicity was observed by mini-beam irradiation. Published studies, however, mostly use rigid mini-beam collimators dedicated to their exact experimental arrangement such that changing the setup or testing new mini-beam collimator configurations becomes challenging and expensive.
Purpose: In this work, a versatile, low-cost mini-beam collimator was designed and manufactured for pre-clinical applications with X-ray beams. The mini-beam collimator enables variability of the full width at half maximum (FWHM), the center-to-center distance (ctc), the peak-to-valley dose ratio (PVDR), and the source-to-collimator distance (SCD).
Methods: The mini-beam collimator is an in-house development, which was constructed of 10 × 40 mm tungsten or brass plates. These metal plates were combined with 3D-printed plastic plates that can be stacked together in the desired order. A standard X-ray source was used for the dosimetric characterization of four different configurations of the collimator, including a combination of plastic plates of 0.5, 1, or 2 mm width, assembled with 1 or 2 mm thick metal plates. Irradiations were done at three different SCDs for characterizing the performance of the collimator. For the SCDs closer to the radiation source, the plastic plates were 3D-printed with a dedicated angle to compensate for the X-ray beam divergence, making it possible to study ultra-high dose rates of around 40 Gy/s. All dosimetric quantifications were performed using EBT-XD films. Additionally, in vitro studies with H460 cells were carried out.
Results: Characteristic mini-beam dose distributions were obtained with the developed collimator using a conventional X-ray source. With the exchangeable 3D-printed plates, FWHM and ctc from 0.52 to 2.11 mm, and from 1.77 to 4.61 mm were achieved, with uncertainties ranging from 0.01% to 8.98%, respectively. The FWHM and ctc obtained with the EBT-XD films are in agreement with the design of each mini-beam collimator configuration. For dose rates in the order of several Gy/min, the highest PVDR of 10.09 ± 1.08 was achieved with a collimator configuration of 0.5 mm thick plastic plates and 2 mm thick metal plates. Exchanging the tungsten plates with the lower-density metal brass reduced the PVDR by approximately 50%. Also, increasing the dose rate to ultra-high dose rates was feasible with the mini-beam collimator, where a PVDR of 24.26 ± 2.10 was achieved. Finally, it was possible to deliver and quantify mini-beam dose distribution patterns in vitro.
Conclusions: With the developed collimator, we achieved various mini-beam dose distributions that can be adjusted according to the needs of the user in regards to FWHM, ctc, PVDR and SCD, while accounting for beam divergence. Therefore, the designed mini-beam collimator may enable low-cost and versatile pre-clinical research on mini-beam irradiation.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/mp.16432 | DOI Listing |
Front Oncol
July 2024
Department of Radiation Oncology, University of Kansas Medical Center, Kansas City, KS, United States.
Background: Proton minibeam radiation therapy (pMBRT) can deliver spatially fractionated dose distributions with submillimeter resolution. These dose distributions exhibit significant heterogeneity in both depth and lateral directions. Accurate characterization of pMBRT doses requires dosimetry devices with high spatial resolution and a wide dynamic range.
View Article and Find Full Text PDFMed Phys
June 2024
Department of Radiation Oncology, University of Kansas Medical Center, Kansas City, Kansas, USA.
Background: Minibeam represents a preclinical spatially fractionated radiotherapy modality with great translational potential. The advantage lies in its high therapeutic index (compared to GRID and LATTICE) and ability to treat at greater depth (compared to microbeam). Proton minibeam radiotherapy (pMBRT) is a synergy of proton and minibeam.
View Article and Find Full Text PDFMed Phys
August 2024
Department for Physics and Astronomy, Heidelberg University, Heidelberg, Germany.
Background: Particle mini-beam therapy exhibits promise in sparing healthy tissue through spatial fractionation, particularly notable for heavy ions, further enhancing the already favorable differential biological effectiveness at both target and entrance regions. However, breathing-induced organ motion affects particle mini-beam irradiation schemes since the organ displacements exceed the mini-beam structure dimensions, decreasing the advantages of spatial fractionation.
Purpose: In this study, the impact of breathing-induced organ motion on the dose distribution was examined at the target and organs at risk(OARs) during carbon ion mini-beam irradiation for pancreatic cancer.
Med Phys
August 2023
Biomedical Physics in Radiation Oncology, German Cancer Research Center (DKFZ), Heidelberg, Germany.
Background: Interest in spatial fractionation radiotherapy has exponentially increased over the last decade as a significant reduction of healthy tissue toxicity was observed by mini-beam irradiation. Published studies, however, mostly use rigid mini-beam collimators dedicated to their exact experimental arrangement such that changing the setup or testing new mini-beam collimator configurations becomes challenging and expensive.
Purpose: In this work, a versatile, low-cost mini-beam collimator was designed and manufactured for pre-clinical applications with X-ray beams.
JCO The purpose of this update was to determine differences in patient-reported chronic toxicity and disease outcomes with intensity-modulated radiation therapy (IMRT) compared with conventional pelvic radiation. Patients with cervical and endometrial cancers who received postoperative pelvic radiation were randomly assigned to conventional radiation therapy (CRT) or IMRT. Toxicity and quality of life were assessed using Patient-Reported Outcomes version of the Common Terminology Criteria for Adverse Events, Expanded Prostate Cancer Index Composite (EPIC) bowel and urinary domains, and Functional Assessment of Cancer Therapy-General.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!