Objective: Zygomatic implant insertion surgery is a challenging operation. The primary aim of this pilot study was to assess the accuracy of EZgoma® "Inverted Support Technique" for the zygomatic implant-guided surgery. Secondly, any factors which may affect the surgical protocol results, such as implant-prosthetic virtual plan, surgical model matching, intra or post-operative complications, time rate between surgical procedure and prosthetic loading, zygomatic implant survival rate and implant success rate were analyzed.
Patients And Methods: A total of 20 zygomatic implants were placed in atrophic maxillae of 5 patients. The final position of zygomatic implants after surgery was compared with the pre-operative digitally planned position. The analyzed parameters were zygomatic implants apex and base mean linear distance and zygomatic implants axis mean angular deviation.
Results: The comparison was provided by a tridimensional imaging elaboration platform, provided by Geomagic, which allows the overlay of virtual plan STL data with post-operative control CT scan DICOM data. As a result, all the mean values regarding the 20 placed zygomatic implants respected the universally agreed values in guided zygomatic implant surgery: the mean linear distance of the implant platform and of the implant apex were 1.59 mm and 1.62 mm respectively, while the mean angular deviation of the implant axis was equal to 1.74°. One of the patients had mucositis as a post-operative complication. In one patient the anterior wall of the maxillary sinus fractured, and in one zygomatic implant primary stabilization was not achieved. No other complications occurred.
Conclusions: As a conclusion, data obtained from this study suggested that guided zygomatic implant rehabilitation may represent a reliable, efficient, rapid, ergonomic, and safe surgical protocol, however further investigations are needed.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.26355/eurrev_202304_31324 | DOI Listing |
J Oral Implantol
January 2025
PhD. Department of Periodontics, Saveetha Dental College, SIMATS, Saveetha University, Chennai, Tamil Nadu, India.
J Craniomaxillofac Surg
December 2024
School of Medicine and Health Sciences, University of Barcelona, Barcelona, Spain; IDIBELL Institute, Barcelona, Spain.
Real-time surgical navigation systems (dynamic computer-aided surgery, d-CAIS) and static guided surgery (static computer-aided surgery, s-CAIS) have been shown to enhance the accuracy of zygomatic implant (ZI) placement. The objective of this systematic review was to evaluate and compare the accuracy and risk of complications associated with d-CAIS and s-CAIS in ZI placement. A systematic review of published studies involving more than 4 patients was conducted to assess and compare the accuracy of d-CAIS and s-CAIS in zygomatic implant placement.
View Article and Find Full Text PDFJ Stomatol Oral Maxillofac Surg
December 2024
Face Ahead® Surgicenter, Belgium and Ziekenhuis aan de Stroom, Campus GZA, B-2018, Antwerp, Belgium. Electronic address:
Objective: This expert opinion presents provisional guidelines for addressing complications associated with Additively Manufactured Subperiosteal Jaw Implants (AMSJI®) in patients with severe maxillary atrophy. AMSJI®'s custom design, supported by finite element analysis (FEA), allows precise placement that avoids critical anatomical structures and minimizes complications relative to alternative solutions.
Materials And Methods: Data were gathered through firsthand experiences, direct communications, and insights from international workgroup meetings.
BMC Oral Health
December 2024
Department of Orthodontics, Institute of Craniofacial Deformity, Yonsei University College of Dentistry, 50-1 Yonsei-Ro, Seodaemun-Gu, Seoul, 03722, Republic of Korea.
Background: This case report details a true hemifacial hyperplasia with temporomandibular joint ankylosis case managed through a multidisciplinary approach involving department of oral and maxillofacial surgery, orthodontics, and prosthodontics.
Case Presentation: A 42-year-old female patient presented with a chief complaint of limited mouth opening. Clinically, the patient exhibited severe facial asymmetry due to hyperplasia of the left facial region.
PeerJ
December 2024
Department of Oral and Maxillofacial Surgery, University Medical Center Groningen, Groningen, Netherlands.
Purpose: Placement of zygomatic implants in the most optimal prosthetic position is considered challenging due to limited bone mass of the zygoma, limited visibility, length of the drilling path and proximity to critical anatomical structures. Augmented reality (AR) navigation can eliminate some of the disadvantages of surgical guides and conventional surgical navigation, while potentially improving accuracy. In this human cadaver study, we evaluated a developed AR navigation approach for placement of zygomatic implants after total maxillectomy.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!