Purpose: The aim was to assess the mid-term aortic remodeling and bare-metal stent (BMS) integrity of the restricted bare stent (RBS) technique reconstruction in aortic dissections.

Materials And Methods: This retrospective cohort study included prospectively collected patients treated with the modified RBS technique between 2017 and 2020. The preoperative, postoperative, and last follow-up computed tomographic (CT) scans were analyzed in the centerline at the mid-descending, celiac trunk (CeT), and the mid-abdominal levels for false lumen (FL) patency, aortic diameter, and true lumen (TL) diameter changes. Bare-metal stent integrity was assessed in the 3-dimensional multiplanar reformats.

Results: The median follow-up of the cohort (n=17) was 26 (11, 45) months. The procedure was mainly performed with the Relay NBS endograft (15/17; 88%) + E-XL BMS (17/17; 100%). Postoperative mortality, paraplegia, stroke, renovisceral vessel loss, and type I and III endoleaks were not observed. BMS fractured in 6 patients (6/17; 36%), damaged the dissection flap in 4/17 (24%), and led to the reperfusion of the FL and re-interventions with TEVAR (4/17; 24%). Two patients without FL reperfusion showed stable CT follow-ups 13 and 17 months after the fracture diagnosis. The TL expansion was seen at all landmarks and peaked in the thoracic aorta (+10; 6, 15; p<0.001). The FL thrombosis after modified RBS was only relevant in the thoracic aorta (p<0.001) and at CeT (p=0.003). The aortic diameter was stable in the thoracic aorta and increased at distal landmarks (CeT [+5; 1, 10; p=0.001]; mid-abdominal [+3; 1, 5; p=0.004]).

Conclusion: The modified RBS technique could not stop aortic growth below the diaphragm and prevent new membrane rupture due to the fractures of the BMS and consecutive flap damage with the reperfusion of the FL.

Clinical Impact: The treatment of complicated type B aortic dissections with TEVAR has become a standard. Particularly, patients with true lumen collapse and malperfusion may benefit from a more aggressive treatment strategy including proximal TEVAR and distal bare-metal stent implantation to re-open the true lumen and to prevent distal stent-induced new entry. However, this study reports the challenges of this approach with a high rate of bare-metal stent fractures during the follow-up. The fractures that occurred at the site of vertical nitinol bridges led to the dissection membrane ruptures and the reperfusion of the false lumen with consecutive dilatation. A close follow-up is mandatory to detect this complication and to treat the patients with TEVAR extension.

Download full-text PDF

Source
http://dx.doi.org/10.1177/15266028231170114DOI Listing

Publication Analysis

Top Keywords

bare stent
12
rbs technique
12
stent rbs
8
bare-metal stent
8
4/17 24%
8
stent fracture
4
fracture tevar
4
tevar modified
4
modified restrictive
4
restrictive bare
4

Similar Publications

Percutaneous Coronary Intervention (PCI) is a treatment method that involves reopening narrowed arteries with a balloon catheter that delivers a cylindrical, mesh-shaped implant device to the site of the stenosis. Currently, by applying a coating to a bare metal stent (BMS) surface to improve biocompatibility, the main risks after PCI, such as restenosis and thrombosis, are reduced while maintaining the basic requirements for the mechanical behavior of the stent itself. In this work, for the first time, the development and optimization process of the spatial structure of the Co-Cr stent (L-605) with a graphene-based coating using cold-wall chemical vapor deposition (CW-CVD) to ensure uniform coverage of the implant was attempted.

View Article and Find Full Text PDF
Article Synopsis
  • The study evaluates the safety and effectiveness of the Zenith Dissection Endovascular System (ZDES) against traditional stent grafts for repairing acute complicated Type B Aortic Dissection (AcTBAD).
  • A retrospective analysis of 32 patients revealed that the ZDES group experienced fewer postoperative adverse events (12 vs. 37), demonstrating better outcomes in false lumen thrombosis at 6 months and lower 3-year mortality rates.
  • Despite some patients needing secondary interventions in the ZDES cohort due to aortic growth, overall results suggest the ZDES may offer improved safety and efficacy compared to traditional methods.
View Article and Find Full Text PDF

Intraoperative rescue of a dislodged renal stent during fenestrated endovascular aortic repair for treatment of type 1A endoleak.

J Vasc Surg Cases Innov Tech

February 2025

Division of Vascular and Endovascular Surgery, Cardio-Thoracic-Vascular Department, Integrated University Healthcare Giuliano-Isontina, University Hospital of Cattinara, Trieste, Italy.

In the past 15 years, fenestrated-branched endovascular aortic repair (F-BEVAR) has progressively become the first-line option for management of most complex abdominal aortic aneurysms (AAAs); with increasing experience, as well as persistent technological refinements, F-BEVAR indications have been expanded to include rescue of failures after prior EVAR. Despite the feasibility and effectiveness, F-BEVAR procedures in the presence of prior infrarenal endografts may come with higher technical complexity that should be properly anticipated, and several anatomical challenges can be expected. Among these, presence of suprarenal bare stents from prior EVAR device are certainly a frequent scenario and may sometimes make target vessel cannulation more difficult because of encroachment on the target vessel origins.

View Article and Find Full Text PDF

Baicalin-based composite coating for achieving biological multifunctionality of vascular stents.

J Control Release

January 2025

Institute of Biomedical Engineering, College of Medicine, Southwest Jiaotong University, Chengdu 610031, PR China. Electronic address:

Article Synopsis
  • Vascular stents are crucial for treating atherosclerosis and have reduced mortality through bare metal and drug-eluting options, but concerns like restenosis and thrombus formation persist.
  • Hyaluronic acid (HA) is often used to modify stent surfaces, but its effectiveness is limited due to degradation by reactive oxygen species (ROS) and enzymes, which can provoke inflammation and affect stent performance.
  • The study introduces a new coating using baicalin (BCL) to protect HA from degradation, offering enhanced endothelial cell support and greater stability, positioning this method as a promising advancement in stent technology.
View Article and Find Full Text PDF

Background: Coronary stenting operations have become the main option for the treatment of coronary heart disease. Vessel recovery after stenting has emerged as a critical factor in reducing possible complications. In this study, we evaluated the feasibility, safety and efficacy of locally administered intraluminal gene therapy delivered using a specialized infusion balloon catheter.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!