2D IMRT QA passing rate dependency on coronal plane.

Phys Med

Department of Radiation Oncology, Fox Chase Cancer Center, Philadelphia, PA 19111, USA.

Published: June 2023

Intensity modulated radiation therapy (IMRT) delivery involves a complex series of beam angles and multileaf collimator (MLC) arrangements, requiring quality assurance to be performed to validate delivery before treatment. The purpose of this work is to investigate the effect of dose gradient on quality assurance (QA) passing rate. Many (n = 40) IMRT plans were delivered and measured using a 2D planar array of ion chambers; additionally, eleven plans were measured at several coronal planes. For each measurement, dose gradient was assessed using a number of metrics and passing rate assessed at both 3%/3 mm and 3%/2 mm criteria. The passing rates of the various IMRT plans were shown to be generally correlated to gradient, with an average distance correlation of 0.54 ± 0.04 for the lateral dose gradient. The passing rate for an individual plan was shown to vary with coronal slice, though the correlation to dose gradient was not predictable. Even though the passing rate was strongly related to dose gradient for many of the plans, the signs of the correlations were not always negative, as hypothesized. The coronal plane at which QA is performed affects passing rate, though dose gradient may not easily be used to predict slices at which passing rate is higher.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ejmp.2023.102594DOI Listing

Publication Analysis

Top Keywords

passing rate
28
dose gradient
24
coronal plane
8
quality assurance
8
imrt plans
8
rate dose
8
rate
7
gradient
7
passing
7
dose
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!