Background: A new keratometric routine that employs power vector management for manual keratometers is described. This study evaluates the agreement of the new proposed keratometric technique with the classical one.
Research Design And Methods: The applicability of a new keratometric routine was verified using Helmholtz's and Javal's keratometers. Results were obtained by two different and well-trained examiners over two different samples, one including 65 and the other 74 eyes, respectively. Both conventional keratometry and the newly proposed routine (named vecto-keratometry) were used in each eye to obtain the results. The clinical agreement between the methods was evaluated using Bland-Altman and Passing-Bablok analysis.
Results: For Helmholtz's keratometer, Bland-Altman plots showed good agreement between methods for both astigmatic components being J = 0.04 ± 0.20 D and J = -0.07 ± 0.17 D. For Javal's keratometer, Passing-Bablok regression test determined regression line for J difference as y = 1.03, confidence interval: [0.98, 1.10] and regression line for J difference as y = 0.97, confidence interval: [0.83, 1.12].
Conclusions: Vecto-keratometry provides accurate clinical results. It has been demonstrated that there are no significant differences between methods in any of the power vector astigmatic components; thus, both methods can be applied interchangeably.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1080/17434440.2023.2206019 | DOI Listing |
Expert Rev Med Devices
June 2023
Department of Optics and Optometry and Vision Sciences, University of Valencia, Burjassot, Spain.
Background: A new keratometric routine that employs power vector management for manual keratometers is described. This study evaluates the agreement of the new proposed keratometric technique with the classical one.
Research Design And Methods: The applicability of a new keratometric routine was verified using Helmholtz's and Javal's keratometers.
Vet Ophthalmol
March 2023
Department of Biostatistics, Bioscience Institute, São Paulo State University - UNESP, São Paulo, Brazil.
Objective: The aims of this study were to compare the central corneal thickness (CCT) measured by an ultrasonic pachymeter (SP-100, Tomey, Japão) versus the Galilei G6™ (Ziemer Ophthalmics System AG, Port, Switzerland) and to compare the simulated keratometry (SimK) measured by an autokeratometer (KM 500 Nidek) versus the Galilei G6™.
Animal Studied: Eighteen mixed-breed cats, 10 males, and eight females, aged between 18 and 48 months, were evaluated (n = 36 eyes).
Procedure(s): The cats were manually restrained using a blanket.
J Curr Ophthalmol
January 2022
Health Promotion Research Center, Zahedan University of Medical Sciences, Zahedan, Iran.
Purpose: To simplify the fitting process, this study was designed to predict the initial power of contact lenses using the regression model based on manual Javal keratometry data and refractive errors.
Methods: In this retrospective study, 121 eyes of 69 patients with keratoconus (KCN) were fitted with a specific trial set of rigid gas permeable contact lenses based on the standard criterion of "three-point touch" over a 7-year period. Power of the cornea was measured using Javal keratometer.
Ophthalmic Physiol Opt
September 2021
Department of Optometry and Vision Science, School of Allied Health Sciences, College of Health and Allied Sciences, University of Cape Coast, Cape Coast, Ghana.