Purpose: With online adaptive radiotherapy (ART), patient-specific quality assurance (PSQA) testing cannot be performed prior to delivery of the adapted treatment plan. Consequently, the dose delivery accuracy of adapted plans (i.e., the ability of the system to interpret and deliver the treatment as planned) are not initially verified. We investigated the variation in dose delivery accuracy of ART on the MRIdian 0.35 T MR-linac (Viewray Inc., Oakwood, USA) between initial plans and their respective adapted plans, by analyzing PSQA results.

Methods: We considered the two main digestive localizations treated with ART (liver and pancreas). A total of 124 PSQA results acquired with the ArcCHECK (Sun Nuclear Corporation, Melbourne, USA) multidetector system were analyzed. PSQA result variations between the initial plans and their respective adapted plans were statistically investigated and compared with the variation in MU number.

Results: For the liver, limited deterioration in PSQA results was observed, and was within the limits of clinical tolerance (Initial = 98.2%, Adapted = 98.2%, p = 0.4503). For pancreas plans, only a few significant deteriorations extending beyond the limits of clinical tolerance were observed and were due to specific, complex anatomical configurations (Initial = 97.3%, Adapted = 96.5%, p = 0.0721). In parallel, we observed an influence of the increase in MU number on the PSQA results.

Conclusion: We show that the dose delivery accuracy of adapted plans, in terms of PSQA results, is preserved in ART processes on the 0.35 T MR-linac. Respecting good practices, and minimizing the increase in MU number can help to preserve the accuracy of delivery of adapted plans as compared to their respective initial plans.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10402677PMC
http://dx.doi.org/10.1002/acm2.14005DOI Listing

Publication Analysis

Top Keywords

adapted plans
20
dose delivery
16
delivery accuracy
16
initial plans
12
plans
9
online adaptive
8
adaptive radiotherapy
8
delivery adapted
8
accuracy adapted
8
035 t mr-linac
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!