Rescue retrieval of a trapped embolic protection device using a JB2-type balloon catheter.

Radiol Case Rep

Department of Neurosurgery, Nakamura Memorial Hospital, South 1, West 14, Chuo-ku, Sapporo, Hokkaido 060-8570, Japan.

Published: June 2023

Carotid artery stenting is a treatment option for patients with significant symptomatic or asymptomatic carotid artery stenosis. Although use of an embolic protection device during carotid artery stenting can prevent distal embolization, device retrieval after stenting can be difficult. In this report, during an embolic protection device retrieval, it was not possible to advance recapture catheters through the distal flare segment of the stent because the device wire would catch on the flare or become jammed between the flare and vessel wall. Then, an extension wire was connected to the end of the device and a 5.2-Fr JB2-type balloon catheter was advanced over the wire, which was able to pass through the stent. The use of a 5.2-Fr JB2-type balloon catheter to retrieve a trapped embolic protection device is feasible and effective if standard maneuvers fail.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10113761PMC
http://dx.doi.org/10.1016/j.radcr.2023.02.059DOI Listing

Publication Analysis

Top Keywords

embolic protection
16
protection device
16
jb2-type balloon
12
balloon catheter
12
carotid artery
12
trapped embolic
8
artery stenting
8
device retrieval
8
52-fr jb2-type
8
device
7

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!