Purpose: A photon Monte Carlo (MC) model was commissioned for flattened (FF) and flattening filter free (FFF) 6 MV beam energy. The accuracy of this model, as a single model to be used for three beam matched LINACs, was evaluated.

Methods: Multiple models were created in RayStation v.10A for three linacs equipped with Elekta "Agility" collimator. A clinically commissioned collapsed cone (CC) algorithm (GoldCC), a MC model automatically created from the CC algorithm without further optimization (CCtoMC) and an optimized MC model (GoldMC) were compared with measurements. The validation of the model was performed by following the recommendations of IAEA TRS 430 and comprised of basic validation in a water tank, validation in a heterogeneous phantom and validation of complex IMRT/VMAT paradigms using gamma analysis of calculated and measured dose maps in a 2D-Array.

Results: Dose calculation with the GoldMC model resulted in a confidence level of 3% for point measurements in water tank and heterogeneous phantom for measurements performed in all three linacs. The same confidence level resulted for GoldCC model. Dose maps presented an agreement for all models on par to each other with γ criteria 2%/2mm.

Conclusions: The GoldMC model showed a good agreement with measured data and is determined to be accurate for clinical use for all three linacs in this study.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ejmp.2023.102588DOI Listing

Publication Analysis

Top Keywords

three linacs
12
model
9
monte carlo
8
dose calculation
8
goldcc model
8
water tank
8
heterogeneous phantom
8
dose maps
8
goldmc model
8
confidence level
8

Similar Publications

On the correction factors for small field dosimetry in 1.5T MR-linacs.

Phys Med Biol

January 2025

Medical Physics Laboratory, School of Medicine, National and Kapodistrian University of Athens, 75 Mikras Asias, Goudi, Athens, Attica, 11527, GREECE.

Clinical dosimetry in the presence of a 1.5T magnetic field is challenging, let alone in case small fields are involved. The scope of this study is to determine a set of relevant correction factors for a variety of MR-compatible detectors with emphasis on small fields.

View Article and Find Full Text PDF

. Adult medulloblastoma (AMB) patients should receive postoperative craniospinal irradiation (CSI) as a standard treatment. Volumetric intensity-modulated arc therapy (VMAT) is a promising method for CSI.

View Article and Find Full Text PDF

Advancements in radiotherapy technology now enable the delivery of ablative doses to targets in the upper urinary tract, including primary renal cell carcinoma (RCC) or upper tract urothelial carcinomas (UTUC), and secondary involvement by other histologies. Magnetic resonance imaging-guided linear accelerators (MR-Linacs) have shown promise to further improve the precision and adaptability of stereotactic body radiotherapy (SBRT). This single-institution retrospective study analyzed 34 patients (31 with upper urinary tract non-metastatic primaries [RCC or UTUC] and 3 with metastases of non-genitourinary histology) who received SBRT from August 2020 through September 2024 using a 1.

View Article and Find Full Text PDF

Background And Purpose: Patients with cardiac implantable electronic devices (CIED patients) are often ineligible for online magnetic resonance-guided radiotherapy (MRgRT), most likely due to the absence of established guidelines. Existing radiotherapy (RT) and magnetic resonance imaging (MRI) guidelines offer an opportunity to construct MRgRT protocols, promoting equitable access. Our objective was to present such a workflow, share multi-institutional experiences treating CIED patients with MRgRT on a 1.

View Article and Find Full Text PDF

Arteriovenous malformations treated by stereotactic radiosurgery - Review of an Australian single centre's experience.

J Clin Neurosci

December 2024

Department of Radiation Oncology, Nelune Comprehensive Cancer Centre, Prince of Wales Hospital, Randwick, 2031 New South Wales, Australia; School of Clinical Medicine (Randwick Campus), Faculty of Medicine and Health, University of New South Wales, Randwick, 2031 New South Wales, Australia; Department of Radiation Oncology, Tamworth Base Hospital, Tamworth, 2340 New South Wales, Australia.

Linear accelerator (LINAC) stereotactic radiosurgery (SRS) is a specialised treatment used for various brain conditions, including arteriovenous malformations (AVMs). This paper investigates the obliteration rate of AVMs treated by the LINAC-based facility at Prince of Wales Hospital, defines factors influencing obliteration rate, time to obliteration and complications post-SRS. A retrospective audit review of patient notes sourced from electronic medical records was conducted.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!