Background: Harmonization methods reduce variability between different make and models of positron emission tomography (PET) scanners. The study aims to explore harmonization strategies that lead to comparable and robust quantitative metrics in a multicenter setting.

Methods: NEMA IEC Phantom data acquisition was performed for low and high spheres-to-background ratios (SBR4:1 and 10:1) on six PET/CT (computed tomography) scanners. Different reconstruction sets, including the number of sub-iterations, number of subsets, and full width at half maximum (FWHM) for each scanner, were evaluated towards optimized and harmonized reconstruction settings. Recovery coefficients (RCs) of four quantitative metrics, including standardized uptake value (SUV), SUV (SUV in 50% isocontour), SUV, and mean uptake of 10 highest concentration voxels were evaluated as RC, RC, RC, and RC, representing percent difference relative to the static ground truth case as functions of sphere sizes. A set of image reconstruction parameters was proposed for harmonized reconstruction to minimize variability between scanners. The root mean square error (RMSE), curvature, and reproducibility were examined. The proposed reconstruction protocols for harmonization and standard clinical reconstruction settings were compared to each other across all scanners.

Results: A significant difference (P value <0.0001) was observed in the aforementioned quantitative metrics between SBR10 and SBR4. Reconstruction parameter sets with the smallest RMSE and RC values within 10% bias were identified as the best candidate for harmonization. The coefficient of variation of the mean value of RCs (CV) shows a remarkable reduction of about 28%, 26%, 32%, and 19% in harmonized reconstruction settings for MRC, MRC, MRC, and MRC, respectively. CV for MRC in the harmonized reconstruction setting was 5.9% in SBR4, while the smallest value in SBR10 belongs to MRC with a value of 5.8%. The reproducibility of RC is improved by deriving the value from ten hottest voxels and is equally reproducible with RC. Compared to RC and RC, the variability is reduced by 18% and 22% if ten voxels are pooled.

Conclusions: Harmonizing PET/CT systems with and without point spread function/time of flight (PSF/TOF) using various vendor-developed image reconstruction algorithms improves the quantification reproducibility. RC, likewise RC, is superior to the rest of the quantitative indices in terms of accuracy and reproducibility and helpful in quantifying lesion volume below 1 mL.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10102741PMC
http://dx.doi.org/10.21037/qims-22-443DOI Listing

Publication Analysis

Top Keywords

positron emission
8
emission tomography
8
quantitative metrics
8
harmonized reconstruction
8
reconstruction settings
8
suv suv
8
reconstruction
6
multicenter quantitative
4
quantitative f-fluorodeoxyglucose
4
f-fluorodeoxyglucose positron
4

Similar Publications

Remote, digital cognitive testing on an individual's own device provides the opportunity to deploy previously understudied but promising cognitive paradigms in preclinical Alzheimer's disease (AD). The Boston Remote Assessment for NeuroCognitive Health (BRANCH) captures a personalized learning curve for the same information presented over seven consecutive days. Here, we examined BRANCH multi-day learning curves (MDLCs) in 167 cognitively unimpaired older adults (age = 74.

View Article and Find Full Text PDF

Anti-amyloid therapy and cerebral blood flow changes on Magnetic Resonance Imaging: a potential longitudinal biomarker of treatment response?

AJNR Am J Neuroradiol

January 2025

From the Department of Department of Radiology, Brain Health Imaging Institute (A.R-F, J.I, S.P, M.d, G.C.C) Weill Cornell Medicine, New York-Presbyterian Hospital, New York, New York, USA; the Department of Neurology (A.R-F), Pontificia Universidad Javeriana, Bogota, Colombia; the Department of Radiology, Division of Molecular Imaging and Therapeutics (A.R-F, J.I) Weill Cornell Medicine, New York-Presbyterian Hospital, New York, New York, USA; the Department of Neurology (D.Z, MM, L.R, A.S.N) Weill Cornell Medicine, New York-Presbyterian Hospital, New York, New York, USA.

Amyloid-targeting therapy has recently become widely available in the U.S. for the treatment of patients with symptomatic mild Alzheimer's disease (AD).

View Article and Find Full Text PDF

Importance: Although 18F-fluorodeoxyglucose positron emission tomography (FDG-PET) is a well-established cross-sectional biomarker of brain metabolism in dementia with Lewy bodies (DLB), the longitudinal change in FDG-PET has not been characterized.

Objective: To investigate longitudinal FDG-PET in prodromal DLB and DLB, including a subsample with autopsy data, and report estimated sample sizes for a hypothetical clinical trial in DLB.

Design, Setting, And Participants: Longitudinal case-control study with mean (SD) follow-up of 3.

View Article and Find Full Text PDF

Coronary artery calcification (CAC) is a key marker of coronary artery disease (CAD) but is often underreported in cancer patients undergoing non-gated CT or PET/CT scans. Traditional CAC assessment requires gated CT scans, leading to increased radiation exposure and the need for specialized personnel. This study aims to develop an artificial intelligence (AI) method to automatically detect CAC from non-gated, freely-breathing, low-dose CT images obtained from positron emission tomography/computed tomography scans.

View Article and Find Full Text PDF

Objective: To explore whether the inflammatory activity is higher in white matter (WM) tracts disrupted by paramagnetic rim lesions (PRLs) and if inflammation in PRL-disrupted WM tracts is associated with disability in people with multiple sclerosis (MS).

Methods: Forty-four MS patients and 16 healthy controls were included. 18 kDa-translocator protein positron emission tomography (TSPO-PET) with the C-PK11195 radioligand was used to measure the neuroinflammatory activity.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!