Background: Intervertebral disc nucleus degeneration initiates a degenerative cascade and can induce chronic low back pain. Nucleus replacement aims to replace the nucleus while the annulus is still intact. Over time, several designs have been introduced, but the definitive solution continues to be elusive. Therefore, we aimed to create a new nucleus replacement that replicates intact intervertebral disc biomechanics, and thus has the potential for clinical applications.
Materials And Methods: Two implants with an outer ring and one (D2) with an additional midline strut were compared. Static and fatigue tests were performed with an INSTRON 8874 following the American Society for Testing and Materials F2267-04, F2346-05, 2077-03, D2990-01, and WK4863. Implant stiffness was analyzed at 0-300 N, 500-2000 N, and 2000-6000 N and implant compression at 300 N, 1000 N, 2000 N, and 6000 N. Wear tests were performed following ISO 18192-1:2008 and 18192-2:2010. GNU Octave software was used to calculate movement angles and parameters. The statistical analysis package R was used with the Deducer user interface. Statistically significant differences between the two designs were analyzed with ANOVA, followed by a post hoc analysis.
Results: D1 had better behavior in unconfined compression tests, while D2 showed a "jump." D2 deformed 1 mm more than D1. Sterilized implants were more rigid and deformed less. Both designs showed similar behavior under confined compression and when adding shear. A silicone annulus minimized differences between the designs. Wear under compression fatigue was negligible for D1 but permanent for D2. D1 suffered permanent height deformation but kept its width. D2 suffered less height loss than D1 but underwent a permanent width deformation. Both designs showed excellent responses to compression fatigue with no breaks, cracks, or delamination. At 10 million cycles, D2 showed 3-times higher wear than D1. D1 had better and more homogeneous behavior, and its wear was relatively low. It showed good mechanical endurance under dynamic loading conditions, with excellent response to axial compression fatigue loading without functional failure after long-term testing.
Conclusion: D1 performed better than D2. Further studies in cadaveric specimens, and eventually in a clinical setting, are recommended. Level of evidence 2c.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10090247 | PMC |
http://dx.doi.org/10.1186/s10195-023-00692-9 | DOI Listing |
BMC Oral Health
December 2024
Department of Conservative Dentistry, Faculty of Dentistry, Minia University, Minia, Egypt.
Objectives: To compare the mechanical performance of partially replaced (repaired) intra-coronal restorations to totally replaced ones in root canal-treated teeth.
Methods: Thirty maxillary second premolars were selected according to strict criteria, mounted on moulds, and had mesio-occluso-distal (MOD) cavities prepared. Resin composite restorative material was used to perform the initial restoration, followed by aging procedures using thermo-mechanical cycling fatigue to replicate six months of intraoral aging.
Adv Mater
December 2024
MOE Key Laboratory of Macromolecular Synthesis and Functionalization, International Research Center for X Polymers, Department of Polymer Science and Engineering, Zhejiang University, 38 Zheda Road, Hangzhou, 310027, China.
Lightweight cellular materials with high stiffness and excellent recoverability are critically important in structural engineering applications, but the intrinsic conflict between these two properties presents a significant challenge. Here, a topological cellular hierarchy is presented, designed to fabricate ultra-stiff (>10 MPa modulus) yet super-elastic (>90% recoverable strain) graphene aerogels. This topological cellular hierarchy, composed of massive corrugated pores and nanowalls, is designed to carry high loads through predominantly reversible buckling within the honeycomb framework.
View Article and Find Full Text PDFAdv Mater
December 2024
Laboratory for Multiscale Mechanics and Medical Science, SV LAB, School of Aerospace, Xi'an Jiaotong University, Xi'an, 710049, China.
Despite fatigue free of monolayer graphene, its assemblies, like cellular graphene aerogels (CGA), are usually suffering of frequent fatigue and inherent strength degradation in repeated loading. In this work, by employing multiscale modeling, the highly intrinsic anisotropic mechanical properties of the cell wall due to the layer-by-layer stacked graphene sheets are uncovered, which easily trigger the unique skeleton joints damage during repeated loading and contribute the primary fatigue mechanism of CGA. Conversely, multiscale joint strengthening strategies are proposed by interlayer crosslinking and joint curvation, improving the interlayer interaction, and decreasing interlayer stress during compression, respectively, so as to effectively suppress joint damage to improve fatigue performance of CGA.
View Article and Find Full Text PDFVopr Kurortol Fizioter Lech Fiz Kult
December 2024
National Medical Research Center for Rehabilitation and Balneology, Moscow, Russia.
Unlabelled: Lymphedema is a chronic disabling disease that affects 250 million people worldwide. To this date, it has been proven that treatment of this category of patients should be truly integrated - combining surgical, therapeutic methods and recovery procedures.
Objective: To study the influence of intermittent pneumatic compression (IPC) on microvasculature in patients with lymphedema of the lower extremities.
Int J Biol Macromol
December 2024
School of Chemistry and Chemical Engineering, Yantai University, Shandong Province, 264005, China. Electronic address:
Although the natural antibacterial agent, cinnamaldehyde, has been extensively studied in the field of food packaging, its water solubility and instability limit its further applications. The controllable responsive release can be achieved through encapsulation in responsive emulsion systems based on carboxymethyl chitosan. Herein, a pH-responsive antibacterial emulsion gel was constructed from cinnamaldehyde-loaded oil-in-water emulsion templates.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!