Contrast sensitivity functions measured using simple optics and computer testing.

Ophthalmic Physiol Opt

Department of Health Promotion and Behavior, Institute of Gerontology, University of Georgia, Athens, Georgia, USA.

Published: July 2023

Purpose: Contrast sensitivity function (CSF) testing is a common approach to assessing clinical changes to specific aspects of spatial vision. Different stimulus presentations and testing procedures, however, yield significant differences in CSF curves that are more a feature of the method than the observer. In this study, we designed a simple optical device for measuring CSF that could be directly calibrated and compared with a commonly used computer-based system.

Methods: Twenty-one participants (M = 28.95 ± 10.34 years; 66.7% female; 81.0% non-Hispanic White; best corrected visual acuity 6/9 or better) provided photopic CSFs (from measurements at 1.6, 3.2, 8, 16 and 24 cycles per degree, with spatial frequency presentation randomised) using both the Metropsis test platform and a simple optical device over two test sessions (one session/method, randomised, counterbalanced) separated by 1-7 days. The optical system used 520 nm lasers that were made Lambertian using two integrating spheres with a 3.5° circular exit port. These beams were combined with a beam splitter that allowed constant measurement of light output and contrast modulation using sine-wave gratings on glass. In Metropsis, 2° Gabor stimuli were presented for 0.5 s with either a vertical or a horizontal orientation via a two-alternative forced choice paradigm with contrast modulated until four (first) and eight (last) reversals were complete.

Results: Both methods took approximately the same amount of time to generate a CSF and yielded curves that were consistent with past studies using similar methods but different from each other. The optical system showed a 3.5 times higher maximum sensitivity and yielded higher test-retest reliability.

Conclusions: Using simple optics to measure CSF yields low noise, high sensitivity and reliability. The ability to calibrate the stimuli directly is an advantage over computer-based methods.

Download full-text PDF

Source
http://dx.doi.org/10.1111/opo.13141DOI Listing

Publication Analysis

Top Keywords

contrast sensitivity
8
simple optics
8
simple optical
8
optical device
8
optical system
8
csf
5
contrast
4
sensitivity functions
4
functions measured
4
simple
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!