Article Abstract

Purpose: To evaluate the dependence of the arterial input function (AIF) on the imaging z-axis and its effect on 3D DCE MRI pharmacokinetic parameters as mediated by the SPGR signal equation and Extended Tofts-Kermode model.

Theory: For SPGR-based 3D DCE MRI acquisition of the head and neck, inflow effects within vessels violate the assumptions underlying the SPGR signal model. Errors in the SPGR-based AIF estimate propagate through the Extended Tofts-Kermode model to affect the output pharmacokinetic parameters.

Materials And Methods: 3D DCE-MRI data were acquired for six newly diagnosed HNC patients in a prospective single arm cohort study. AIF were selected within the carotid arteries at each z-axis location. A region of interest (ROI) was placed in normal paravertebral muscle and the Extended Tofts-Kermode model solved for each pixel within the ROI for each AIF. Results were compared to those obtained with a published population average AIF.

Results: Due to inflow effect, the AIF showed extreme variation in their temporal shapes. K was most sensitive to the initial bolus concentration and showed more variation over the muscle ROI with AIF taken from the upstream portion of the carotid. k was less sensitive to the peak bolus concentration and showed less variation for AIF taken from the upstream portion of the carotid.

Conclusion: Inflow effects may introduce an unknown bias to SPGR-based 3D DCE pharmacokinetic parameters. Variation in the computed parameters depends on the selected AIF location. In the context of high flow, measurements may be limited to relative rather than absolute quantitative parameters.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10194023PMC
http://dx.doi.org/10.1016/j.mri.2023.03.016DOI Listing

Publication Analysis

Top Keywords

extended tofts-kermode
12
arterial input
8
input function
8
head neck
8
aif
8
dce mri
8
pharmacokinetic parameters
8
spgr signal
8
spgr-based dce
8
inflow effects
8