An in vitro study evaluating the efficacy of a novel mount with torque control designed to tighten Osstell® transducers.

Clin Exp Dent Res

Department of Restorative Dentistry and Periodontology, Dublin Dental University Hospital, Trinity College, Dublin, Ireland.

Published: June 2023

Objectives: To evaluate the efficacy of a novel mount with torque control for tightening of Osstell® transducers and to determine the reliability of recorded ISQ measurements from implants placed in various bone densities. MATERIAL AND METHODS: Fifty-six implants, comprising seven different implant types, were placed in eight polyurethane blocks representing D1, D2, D3, and D4 bone densities. Resonance frequency analysis (RFA) transducers were attached to each implant in four different ways: (a) hand tightening, (b) hand tightening with a SmartPeg Mount™, (c) hand tightening using the novel mount with torque control (SafeMount) and (d) tightening to 6 Ncm with a calibrated torque device. ISQ measurements were taken and a second operator repeated the measurements. Intraclass correlation coefficient (ICC) was calculated to assess the reliability of the measurements and linear mixed effects regression was employed to determine the effect explanatory variables had on ISQ values.

Results: There was a statistically significant difference in ISQ values obtained by hand tightening transducers compared to the calibrated torque device p < .001, 95%(-2.89, -1.21) but not between any other tightening methods. There was excellent agreement between the two RFA devices (ICC 0.986) and between buccal and mesial measurements (ICC 0.977). For all transducer tightening methods there was excellent inter-operator agreement in D1 and D2 (ICC > 0.8) but very poor agreement in D4 (ICC < 0.24). Bone density accounted for 36% of the variation in ISQ values, the implant for 11% and the operator for 6%.

Conclusions: SafeMount, did not significantly improve the reliability of the RFA measurements when compared to the standard mount, but calibrated torque devices seem to have benefits when compared to tightening the transducers by hand. Results also indicate that the ISQ values should be interpreted with caution when measuring implant stability in poor quality bone regardless of the implant geometry.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10280597PMC
http://dx.doi.org/10.1002/cre2.734DOI Listing

Publication Analysis

Top Keywords

hand tightening
16
novel mount
12
mount torque
12
torque control
12
efficacy novel
8
osstell® transducers
8
isq measurements
8
bone densities
8
calibrated torque
8
torque device
8

Similar Publications

Cosmetically active compounds (CACs), both of lipophilic and hydrophilic origin, have difficulty reaching the deeper layers of the skin, and this shortcoming significantly reduces their efficacy. One such CAC that occurs naturally in the human body and displays many beneficial properties (via reducing fine lines and wrinkles, tightening skin, improving its elasticity, etc.) is the glycyl-L-histidyl-L-lysine tripeptide complex of copper (GHK-Cu).

View Article and Find Full Text PDF

A Novel Sliding Knot Technique Without a Knot Pusher for Laparoscopic Pelvic Floor Surgery.

Int Urogynecol J

November 2024

Koç University School of Medicine, Koç University, Istanbul, Turkey.

Introduction And Hypothesis: Challenges in intracorporeal knot tying can be addressed with extracorporeal slip knots, simplifying the process of tying and managing tissue tension using a knot pusher. However, existing extracorporeal knot techniques are difficult owing to their complexity, the finesse required with thin yarns, and extensive training needs. We developed a new laparoscopic extracorporeal slip knot technique that can be used with a conventional needle driver or standard clamps, offering the advantages of being cost-effective and easy to learn.

View Article and Find Full Text PDF

After failed conservative therapy or in the absence of any intervention, a rupture of the digital subcutaneous extensor tendon at the distal interphalangeal (DIP) joint, known as mallet finger, may lead to a chronic extension deficit due to excessive scarring and tendon elongation. Various surgical techniques to restore the extension of the distal phalanx have been proposed, but an optimal approach has not yet been established. To tighten the extensor tendon, a purse-string suture can be applied.

View Article and Find Full Text PDF

Biomechanical analysis of cadaveric wrists before and after MOTEC wrist arthroplasty using a hexapod robot.

J Hand Surg Eur Vol

September 2024

Biomechanics and Implants Research Group, Medical Device Research Institute, College of Science and Engineering, Flinders University, Bedford Park, South Australia, Australia.

This study compares wrist motion, biomechanical behaviour and radiographic parameters before and after total wrist arthroplasty using a fourth-generation spherical articulation prosthesis. A total of 10 cadaveric specimens were assessed using a hexapod Stewart platform robot. After arthroplasty, there were significant increases in both stiffness and phase angle of wrist motion across all planes of motion assessed.

View Article and Find Full Text PDF

Background And Objectives: Effective microsurgical training necessitates dedicated practice, prompting the development of simulation models that mitigate the hygiene risks, regulatory challenges, and storage difficulties associated with conventional biological models. This study aims to evaluate the preferences of microsurgeons and medical students regarding self-made silicone simulation vessels, comparing them to standard biological models.

Methods: A three-part jig, comprising of a two-part metal clamp component and a transparent acrylic block with 4x4 channels, was designed.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!