Background: The Roche Cobas c Tina-quant C-Reactive Protein (CRP) IV method has shown persistent, significant negative bias compared to other methodologies in the UK NEQAS external quality assurance (EQA) scheme. The aim of this study was to compare CRP concentrations determined by this restandardized 4th generation assay with another commercially available method to assess the extent of any bias and impact on clinical cut-offs.
Methods: CRP concentrations in 38 anonymized patient serum samples were determined by a particle-enhanced immunoturbidimetric assay on the Roche Cobas c701 module and a latex assay on the Beckman Coulter AU5800 analyser over a concentration range of 1-308 mg/L.
Results: 97.4% samples analysed by the Roche Tina-quant CRP IV method demonstrated a negative bias compared to the Beckman AU latex method. The mean absolute bias was -4.12 mg/L (range: -19.13-10.93 mg/L; 95% CI: -17.25-9.01 mg/L). The mean relative difference was -15.5% (range: -40-4%; 95% CI: -33.03-1.94%). This bias was seen across the range of samples assayed, including at clinically significant cut-offs of 5 mg/L (-24% bias), 20 mg/L (-5%) and 100 mg/L (-13%).
Conclusion: The negative bias of the Roche method demonstrated in the EQA scheme appears to reflect genuine differences in patient results, rather than an EQA matrix effect, despite re-standardization of the CRP assay. This report indicates that clinical cut-offs and reference ranges may be method-dependent, which should be reflected in published guidance and interpretation provided by laboratories.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1177/00045632231169151 | DOI Listing |
Clin Infect Dis
January 2025
IQVIA Inc., Falls Church, VA.
Clin Chim Acta
January 2025
Department of Clinical Laboratory, Beijing Chaoyang Hospital, The Third Clinical Medical College of Capital Medical University, Beijing Center for Clinical Laboratories, No. 8 Gongti South Road, Chaoyang District, Beijing 100020, PR China. Electronic address:
Background: Lactate dehydrogenase (LDH) is a critical enzyme widely used in clinical diagnostics. However, variations in measurement systems can lead to inconsistent results, potentially impacting clinical decision-making. This study aimed to evaluate the comparability of lactate dehydrogenase measurements by comparing routine methods with the IFCC reference method, and to analyze the standardization status of LDH testing through external quality assessment (EQA).
View Article and Find Full Text PDFBackground: Serum B-cell maturation antigen (sBCMA) levels have emerged as a potential biomarker for disease monitoring in multiple myeloma (MM) with prognostic value.
Methods: Herein, we evaluated the sBCMA levels in 166 patients with newly diagnosed MM with an Elecsys-based sBCMA assay.
Results: Increased sBCMA levels at diagnosis were correlated with inferior survival outcomes in terms of both progression-free and overall survival.
Scand J Clin Lab Invest
January 2025
Department of Medical Biochemistry, Health Science University, Istanbul Basaksehir Cam and Sakura City Hospital, Istanbul, Turkey.
This study assessed the reliability of Roche Accu-Chek Inform II glucometers in a real-world setting. A retrospective analysis was conducted on 6,695 paired results. Capillary samples were tested using Roche Accu-Chek Inform II glucometers, while venous samples were analyzed using Roche Cobas c503/702 analyzers.
View Article and Find Full Text PDFAnn Clin Lab Sci
November 2024
Department of Laboratory Medicine, Kangbuk Samsung Hospital, Sungkyunkwan University School of Medicine, Seoul, Korea
Objective: Molecular testing for high-risk human papillomavirus (hrHPV) genotypes is important for cervical cancer screening. In this study, we compared the HPV detection rates using real-time PCR in cervical samples collected using two different liquid-based cytology (LBC) kits.
Methods: Cervical swab specimens were prospectively collected using the SurePath and EASYPREP collection kits.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!