Purpose: The purpose of this work is to determine the sensitivity of phase-decorrelation optical coherence tomography (OCT) to protein aggregation associated with cataracts in the ocular lens, as compared to OCT signal intensity.

Methods: Six fresh porcine globes were held at 4°C until cold cataracts developed. As the globes were re-warmed to ambient temperature, reversing the cold cataract, each lens was imaged repeatedly using a conventional OCT system. Throughout each experiment, the internal temperature of the globe was recorded using a needle-mounted thermocouple. OCT scans were acquired, their temporal fluctuations were analyzed, and the rates of decorrelation were spatially mapped. Both decorrelation and intensity were evaluated as a function of recorded temperature.

Results: Both signal decorrelation and intensity were found to change with lens temperature, a surrogate of protein aggregation. However, the relationship between signal intensity and temperature was not consistent across different samples. In contrast, the relationship between decorrelation and temperature was found to be consistent across samples.

Conclusions: In this study, signal decorrelation was shown to be a more repeatable metric for quantification of crystallin protein aggregation in the ocular lens than OCT intensity-based metrics. Thus, OCT signal decorrelation measurements could enable more detailed and sensitive study of methods to prevent cataract formation.

Translational Relevance: This dynamic light scattering-based approach to early cataract assessment can be implemented on existing clinical OCT systems without hardware additions, so it could quickly become part of a clinical study workflow or an indication for use for a pharmaceutical cataract intervention.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10064939PMC
http://dx.doi.org/10.1167/tvst.12.3.25DOI Listing

Publication Analysis

Top Keywords

protein aggregation
12
signal decorrelation
12
phase-decorrelation optical
8
optical coherence
8
coherence tomography
8
ocular lens
8
oct signal
8
decorrelation intensity
8
temperature consistent
8
oct
7

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!