Purpose: To compare dosimetric parameters of brachytherapy (BT) treatment plans made with or without active source positions of the ring/ovoid (R/O) applicator in locally advanced cervical cancer patients.
Material And Methods: Sixty patients with cervical cancer without vaginal involvement were selected for the study, who received intra-cavitary/interstitial BT. For each patient, two plans with and without active source dwell positions in R/O were created, using the same dose-volume constraints. EQD total doses from external beam and BT of target volumes and organs at risk (OARs) between the competing plans were compared.
Results: There was no significant difference in the dose of high-risk clinical target volume (HR-CTV) and gross tumor volume (GTV) between the plans with inactive vs. active R/O. The mean D of intermediate-risk clinical target volume (IR-CTV) was significantly lower with inactive R/O; however, the GEC-ESTRO (EMBRACE II study) and ABS criteria were fulfilled in 96% in both plans. There was no difference in dose homogeneity, but conformity of the plans with inactive R/O was higher. Doses to all OARs were significantly lower in plans without R/O activation. While all the plans without R/O activation fulfilled the recommended dose criteria for OARs, it was less achievable with R/O activation.
Conclusions: Inactivation of R/O applicator results in similar dose coverage of the target volumes with lower doses to all OARs, as activation of R/O in cervix cancer patients when HR-CTV does not extend to R/O applicator. The use of active source positions in R/O shows worse performance regarding the fulfilment of the recommended criteria for OARs.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10034727 | PMC |
http://dx.doi.org/10.5114/jcb.2023.124399 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!