Purpose: Suitable commissioning and quality control (QC) tests for high-dose-rate brachytherapy (HDR-BT) is necessary to ensure dosimetric and geometric accuracy of the treatment. This study aimed to present the methodology of developing a novel multi-purpose QC phantom (AQuA-BT) and examples of its' application in 3D image-based (particularly magnetic resonance imaging [MRI]-based) planning for cervix BT.
Material And Methods: Design criteria led to a phantom with sufficient size waterproof box for dosimetry and capability for inserting other components inside the phantom for: (A) Validating dose calculation algorithms in treatment planning systems (TPSs) using a small-volume ionization chamber; (B) Testing volume calculation accuracy in TPSs for bladder, rectum, and sigmoid organs at risk (OARs) constructed by 3D printing; (C) Quantification of MRI distortions using 17 semi-elliptical plates with 4,317 control points to mimic a realistic female's pelvis size; and (D) Quantification of image distortions and artifacts induced by MRI-compatible applicators using a specific radial fiducial marker. The utility of the phantom was tested in various QC procedures.
Results: The phantom was successfully implemented for examples of intended QC procedures. The maximum deviation between the absorbed doses to water assessed with our phantom and those calculated by SagiPlan TPS was 1.7%. The mean discrepancy in volumes of TPS-calculated OARs was 1.1%. The differences between known distances within the phantom on MR imaging were within 0.7 mm compared with computed tomography.
Conclusions: This phantom is a promising useful tool for dosimetric and geometric quality assurance (QA) in MRI-based cervix BT.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10034728 | PMC |
http://dx.doi.org/10.5114/jcb.2023.125014 | DOI Listing |
Br J Radiol
January 2025
Radiotherapy Physics Dept, Ipswich Hospital, Ipswich, Suffolk, IP45PD, UK.
Objectives: To survey kilovoltage (kV) radiotherapy in the UK, updating a 2016 study, focussing on radiotherapy physics, including equipment quality control (QC) and radiation dosimetry, with information on installed equipment and clinical activity.
Methods: All UK radiotherapy physics departments (n = 68) were invited to complete a comprehensive survey. An analysis of the installed equipment base, patient numbers, clinical activity, QC testing and radiation dosimetry processes were undertaken.
Purpose: With the widespread introduction of dual energy computed tomography (DECT), applications utilizing the spectral information to perform material decomposition became available. Among these, a popular application is to decompose contrast-enhanced CT images into virtual non-contrast (VNC) or virtual non-iodine images and into iodine maps. In 2021, photon-counting CT (PCCT) was introduced, which is another spectral CT modality.
View Article and Find Full Text PDFMed Phys
January 2025
Department of Radiation Oncology, Inha University Hospital, Incheon, Republic of Korea.
Background: High-dose-rate (HDR) brachytherapy using Iridium-192 as a radiation source is widely employed in cancer treatment to deliver concentrated radiation doses while minimizing normal tissue exposure. In this treatment, the precision with which the sealed radioisotope source is delivered significantly impacts clinical outcomes.
Purpose: This study aims to evaluate the feasibility of a new four-dimensional (4D) in vivo source tracking and treatment verification system for HDR brachytherapy using a patient-specific approach.
Med Phys
January 2025
Department of Radiation Oncology, University of Texas Southwestern Medical Center, Dallas, Texas, USA.
Background: Diffusion-weighted (DW) turbo-spin-echo (TSE) imaging offers improved geometric fidelity compared to single-shot echo-planar-imaging (EPI). However, it suffers from low signal-to-noise ratio (SNR) and prolonged acquisition times, thereby restricting its applications in diagnosis and MRI-guided radiotherapy (MRgRT).
Purpose: To develop a joint k-b space reconstruction algorithm for concurrent reconstruction of DW-TSE images and the apparent diffusion coefficient (ADC) map with enhanced image quality and more accurate quantitative measurements.
Phys Med Biol
January 2025
North China Electric Power University - Baoding Campus, North China Electric Power University, Baoding, Hebei Province, P.R.China, Baoding, Hebei, 071003, CHINA.
Objective: The optical absorption properties of biological tissues in photoacoustic tomography are typically quantified by inverting acoustic measurements. Conventional approaches to solving the inverse problem of forward optical models often involve iterative optimization. However, these methods are hindered by several challenges, including high computational demands, the need for regularization, and sensitivity to both the accuracy of the forward model and the completeness of the measurement data.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!