Background: The potential reduction of normal tissue toxicities during FLASH radiotherapy (FLASH-RT) has inspired many efforts to investigate its underlying mechanism and to translate it into the clinic. Such investigations require experimental platforms of FLASH-RT capabilities.
Purpose: To commission and characterize a 250 MeV proton research beamline with a saturated nozzle monitor ionization chamber for proton FLASH-RT small animal experiments.
Methods: A 2D strip ionization chamber array (SICA) with high spatiotemporal resolution was used to measure spot dwell times under various beam currents and to quantify dose rates for various field sizes. An Advanced Markus chamber and a Faraday cup were irradiated with spot-scanned uniform fields and nozzle currents from 50 to 215 nA to investigate dose scaling relations. The SICA detector was set up upstream to establish a correlation between SICA signal and delivered dose at isocenter to serve as an in vivo dosimeter and monitor the delivered dose rate. Two off-the-shelf brass blocks were used as apertures to shape the dose laterally. Dose profiles in 2D were measured with an amorphous silicon detector array at a low current of 2 nA and validated with Gafchromic films EBT-XD at high currents of up to 215 nA.
Results: Spot dwell times become asymptotically constant as a function of the requested beam current at the nozzle of greater than 30 nA due to the saturation of monitor ionization chamber (MIC). With a saturated nozzle MIC, the delivered dose is always greater than the planned dose, but the desired dose can be achieved by scaling the MU of the field. The delivered doses exhibit excellent linearity with with respect to MU, beam current, and the product of MU and beam current. If the total number of spots is less than 100 at a nozzle current of 215 nA, a field-averaged dose rate greater than 40 Gy/s can be achieved. The SICA-based in vivo dosimetry system achieved excellent estimates of the delivered dose with an average (maximum) deviation of 0.02 Gy (0.05 Gy) over a range of delivered doses from 3 to 44 Gy. Using brass aperture blocks reduced the 80%-20% penumbra by 64% from 7.55 to 2.75 mm. The 2D dose profiles measured by the Phoenix detector at 2 nA and the EBT-XD film at 215 nA showed great agreement, with a gamma passing rate of 95.99% using 1 mm/2% criterion.
Conclusion: A 250 MeV proton research beamline was successfully commissioned and characterized. Challenges due to the saturated monitor ionization chamber were mitigated by scaling MU and using an in vivo dosimetry system. A simple aperture system was designed and validated to provide sharp dose fall-off for small animal experiments. This experience can serve as a foundation for other centers interested in implementing FLASH radiotherapy preclinical research, especially those equipped with a similar saturated MIC.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/mp.16364 | DOI Listing |
Phys Med Biol
January 2025
Radiotherapy and Radiation Dosimetry group, National Physical Laboratory, Hampton Road, Middlesex, Teddington, TW11 0LW, UNITED KINGDOM OF GREAT BRITAIN AND NORTHERN IRELAND.
Internationally, reference dosimetry for clinical proton beams largely follows the guidelines published by the International Atomic Energy Agency (IAEA TRS-398 Rev. 1, 2024). This approach yields a relative standard uncertainty of 1.
View Article and Find Full Text PDFAppl Radiat Isot
January 2025
National Institute of Metrology (NIM), Beijing, 100029, China.
A current measurement system named ICMS-TIA (Ionization Current Measurement System based on Transimpedance Amplifier) for rapid and accurate measurement of the ionization current of 4πγ ionization chamber is developed based on an improved high-value resistance I-V conversion method with range switching function. The effective measurement range of the ICMS-TIA is from 50 fA to 50 μA and the response time of the system is less than 2.8 s.
View Article and Find Full Text PDFInt J Part Ther
March 2025
Institute of Medical Physics and Radiation Protection, University of Applied Sciences, Giessen, Germany.
Purpose: The spot size of scanned particle beams is of crucial importance for the correct dose delivery and, therefore, plays a significant role in the quality assurance (QA) of pencil beam scanning ion beam therapy.
Materials And Methods: This study compares 5 detector types-radiochromic film, ionization chamber (IC) array, flat panel detector, multiwire chamber, and IC-for measuring the spot size of proton and carbon ion beams.
Results: Variations of up to 30% were found between detectors, underscoring the impact of detector choice on QA outcomes.
Tech Innov Patient Support Radiat Oncol
March 2025
Graduate School of Biomedical Sciences, Tokushima University, Tokushima, Tokushima 770-8503, Japan.
Purpose: This study aims to compare treatment plans created using RapidPlan and PlanIQ for twelve patients with prostate cancer, focusing on dose uniformity, dose reduction to organs at risk (OARs), plan complexity, and dose verification accuracy. The goal is to identify the tool that demonstrates superior performance in achieving uniform target dose distribution and reducing OAR dose, while ensuring accurate dose verification.
Methods: Dose uniformity in the planning target volume, excluding the rectum, and dose reduction in the OARs (the rectum and bladder) were assessed.
J Neurosci Methods
January 2025
Cognitive Neuroscience Laboratory, German Primate Center - Leibniz Institute for Primate Research, Goettingen, Germany; Faculty of Biology and Psychology, University of Goettingen, Goettingen 37077 Germany.
Background: Neuronal activity is modulated by behavior and cognitive processes. The combination of several neurotransmitter systems, acting directly or indirectly on specific populations of neurons, underlie such modulations. Most studies with non-human primates (NHPs) fail to capture this complexity, partly due to the lack of adequate methods for reliably and simultaneously measuring a broad spectrum of neurotransmitters while the animal engages in behavioral tasks.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!