Background: Concerns remain that thinner highly crosslinked polyethylene (HXLPE) liners in modern total hip arthroplasty (THA) may lead to premature liner-related failures or revision. The aim of this study was to evaluate the effect of liner thickness on survival and revision rates of HXLPE in primary THA after more than 10 years.
Methods: We retrospectively identified 2,565 primary THAs using HXLPE with a mean follow-up of 13 years (range, 11 to 19). Patients were grouped for each mm of polyethylene thickness. Liner thickness varied from 4.9 to 12.7 mm, with one third measuring less than 6 mm. Primary outcomes were reoperation, all-cause revision, and liner-related revision.
Results: The reoperation rate was 7.3%, the all-cause revision rate was 5.6%, and the liner-related revision rate was 0.04%. There was no significant difference in all-cause revision when stratified by liner thickness (P = .286) and liner thickness was not associated with liner-related revision (n = 1). There was a statistically, but not clinically significant difference in mean liner thickness for the cohort that underwent reoperation (7.09 versus 6.89 mm, P = .01) and all-cause revision (7.16 versus 6.89 mm, P = .031).
Conclusion: In our cohort, liner thickness was not associated with all-cause revision-free survival, and there was no clinically significant difference in liner thickness between those patients who did require a reoperation or all-cause revision, and those who did not. There was only 1 liner-related failure in the entire cohort. Our results indicate that using thinner HXLPE liners to maximize femoral head size in THA is a safe practice that does not lead to increased revision rates or liner failure at a mean 13-year follow-up.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.arth.2023.03.026 | DOI Listing |
Objective: To evaluate color masking and relative translucency parameter (RTP) of increasing dentin thicknesses from different resin composites, with or without opacifiers, on a veneer dental preparation and resin disks.
Material And Methods: Artificial darkened lateral incisors with 1mm-thick veneers preparations were used to evaluate color masking of different resinous materials, with or without opacifiers: IPS Empress Direct (ED) with or without ED Opaque; and Essentia (ES) with or without ES Masking Liner. For the RTP test; disc-shaped specimens were performed and evaluated with a spectrophotometer (VITA Easyshade) against black and C4 backgrounds.
Int J Clin Pediatr Dent
October 2024
Department of Conservative Dentistry and Endodontics, Subbaiah Institute of Dental Sciences, Shivamogga, Karnataka, India.
Aim: To compare the microleakage of three bulk-fill composite resins with or without resin-modified glass ionomer cement (RMGIC) liner.
Materials And Methods: A total of 30 maxillary human 1st premolar teeth were selected. Two box preparations were made on the mesial and distal sides.
Arthroplast Today
October 2024
Orthopaedic Surgery, Yale University, Yale New Haven Hospital, New Haven, CT, USA.
Background: Robotic systems for total knee arthroplasty (TKA) may utilize computed tomography three-dimensional modeling and intraoperative ligamentous balancing data to assist surgeons with implant size and position. This study evaluated the effect of such robotic systems on implant selection.
Methods: We reviewed 645 TKAs performed with a single prosthetic design at 2 academic medical centers between 2016 and 2022.
PLoS One
November 2024
Joint & Arthritis Research, Department of Orthopaedic Surgery, Himchan Hospital, Seoul, Republic of Korea.
Purpose: This study aimed to compare the clinical outcomes, mechanical axis, component positioning, leg length discrepancy (LLD), and polyethylene liner thickness between robotic-assisted total knee arthroplasty (TKA) and conventional TKA in patients with failed high tibial osteotomy (HTO).
Methods: A total of 30 patients (30 knees) with failed HTO who underwent TKA using a robot-assisted system between June 2020 and December 2023 were included in this study (robotic group). Additionally, 60 patients (60 knees) with failed HTO who underwent conventional TKA were included as controls (conventional group).
Environ Sci Pollut Res Int
November 2024
Institute of Energy Infrastructure (IEI), Universiti Tenaga Nasional, Jalan Ikram-Uniten, 43000, Kajang, Selangor, Malaysia.
In municipal solid waste (MSW) landfills, biodegradation of the organic MSW fraction results in elevated waste and basal liner temperatures which have the potential to cause the clay component of the basal liner to experience severe moisture loss over time and eventually undergo desiccation cracking. Cracking of the basal liner's clay component would result in an uncontrolled release of contaminants into the surrounding environment and ultimately give rise to a variety of major environmental concerns. Accordingly, this study examined the variation of temperature-moisture profiles along the depth of a compacted clay liner (CCL) exposed to different constant elevated waste temperatures (CETs) in the absence and presence of two heat reduction techniques, respectively.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!