Background: Titanium mesh cranioplasty is often performed after decompressive craniectomy. Spontaneous fracture of the titanium prosthesis is an extremely rare postoperative complication. Here, we report a 10-year-old boy who presented with a spontaneous fracture of titanium mesh without antecedent head trauma.

Case Summary: A 10-year-old boy presented with a 1-wk history of a tender bulge over the left temporo-parieto-occipital scalp. He had undergone a temporo-parieto-occipital titanium mesh cranioplasty 26 mo previously. He denied antecedent head trauma. Computerized tomography disclosed a perpendicular fissure in the titanium mesh, suggesting a diagnosis of spontaneous titanium mesh fracture. He underwent a second temporo-parieto-occipital cranioplasty and made an uneventful recovery. Three-dimensional modeling and finite element analyses were used to explore potential risk factors of titanium mesh fracture.

Conclusion: We report a case of spontaneous fracture of a titanium mesh cranioplasty implant. The current case and literature review indicate that titanium mesh implants should be well-anchored to the base of bony defects to prevent fatigue-induced fractures.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10011976PMC
http://dx.doi.org/10.12998/wjcc.v11.i7.1593DOI Listing

Publication Analysis

Top Keywords

titanium mesh
36
spontaneous fracture
16
fracture titanium
16
mesh cranioplasty
16
titanium
10
mesh
9
cranioplasty implant
8
10-year-old boy
8
boy presented
8
antecedent head
8

Similar Publications

Metallic biomaterials are extensively used in orthopedics and dentistry, either as implants or coatings. In both cases, metal ions come into contact with surrounding tissues causing a particular cell response. Here, we present a biofabricated tissue model, consisting of a hydrogel reinforced with a melt electrowritten mesh, to study the effects of bound and released metal ions on surrounding cells embedded in a hydrogel matrix.

View Article and Find Full Text PDF

Background: Posterior vault distraction osteogenesis (PVDO) is an effective treatment for multisuture and syndromic craniosynostosis. It remains unclear how well the calvarium ossifies after PVDO, especially in older patients when they have plateaued in their cranial growth phase. The purpose of this study is to report outcomes associated with PVDO across a wide range of ages at our institution.

View Article and Find Full Text PDF

Evaluation of sticky bone in guided bone regeneration in the aesthetic area of the anterior teeth: a retrospective study.

Int J Oral Maxillofac Surg

January 2025

Stomatological Hospital of Chongqing Medical University, Chongqing, China; Chongqing Municipal Key Laboratory of Oral Biomedical Engineering of Higher Education, Chongqing, China; Chongqing Key Laboratory of Oral Diseases and Biomedical Sciences, Chongqing, China. Electronic address:

The retrospective study aimed to compare the space-maintaining effects of sticky bone (bone graft matrix enriched with injectable platelet-rich fibrin) and titanium mesh for bone augmentation in the aesthetic zone. Patients who underwent single implant placement and had type 2/4 alveolar bone defects (buccal bone wall loss is >50% of the expected implant length) were screened for inclusion in this study. The labial bone plate width was measured at 1, 3, and 5 mm below the apical implant platform on cone beam computed tomography images taken immediately and 6 months after surgery.

View Article and Find Full Text PDF

Intracranial Hypotension Mechanism and Implant Retention Procedure for Patients With Titanium Mesh Exposure.

Oper Neurosurg (Hagerstown)

January 2025

Department of Neurosurgery, Huashan Hospital, Shanghai Medical College, Fudan University, Shanghai, China.

Background And Objectives: The study aimed to investigate the potential pathogenesis and present an implant retention procedure for patients with titanium mesh exposure after cranioplasty.

Methods: The clinical data were obtained from 26 consecutive cases with titanium mesh exposure who underwent surgical treatment between 2018 and 2023. These patients' medical records, scalp photographs, operative notes, and outcomes were retrospectively analyzed.

View Article and Find Full Text PDF
Article Synopsis
  • Cage subsidence can negatively affect lumbar fusion procedures, with material selection (PEEK vs. 3D-Ti) influencing this issue; the study aims to compare their subsidence rates.
  • The systematic search reviewed 265 patients from three high-quality studies, focusing on cage subsidence and classified subsidence rates using a specific method.
  • Results indicated that 3D-Ti cages have a significantly lower rate of subsidence compared to PEEK cages, with less severe subsidence and better overall performance.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!