Background: The third-generation MitraClip NTR/XTR transcatheter edge-to-edge repair system was introduced to assist in leaflet grasping with the longer clip arms of MitraClip XTR and to improve ease of use with the modified delivery catheter.

Objectives: The EXPAND study evaluated contemporary real-world outcomes in subjects with mitral regurgitation (MR) treated with the third-generation MitraClip NTR/XTR transcatheter edge-to-edge repair system.

Methods: EXPAND is a prospective, multicenter, international, single-arm study that enrolled patients with primary MR and secondary MR at 57 centers. Follow-up was conducted through 12 months. Echocardiograms were analyzed by an echocardiographic core laboratories. Study outcomes included: MR severity, functional capacity measured by New York Heart Association functional class, quality of life measured by Kansas City Cardiomyopathy Questionnaire, heart failure hospitalizations, all-cause mortality.

Results: 1,041 patients were enrolled from April 2018 through March 2019, of which 50.5% had primary or mixed etiology. Implant success was 98.9%; 1.5 ± 0.6 clips were implanted per subject. Significant MR reduction from baseline (≥MR 3+: 56.0%) to 30 days (≤MR 1+:88.8%) was maintained through 1 year (MR ≤1+: 89.2%). A total of 84.5% and 93.0% of subjects in primary MR and secondary MR, respectively, had ≤1+ MR at 1 year. Significant improvements were observed in clinical outcomes (New York Heart Association functional class I/II in 80.3%, +21.6 improvement in Kansas City Cardiomyopathy Questionnaire score) at 1 year. All-cause mortality and heart failure hospitalizations at 1 year were 14.9% and 18.9%, respectively, which was significantly lower than previous studies.

Conclusions: The study demonstrates treatment with the third-generation system resulted in substantial reduction of MR in a contemporary real-world practice, compared with the results of earlier EVEREST and COAPT trials.(The MitraClip® EXPAND Study of the Next Generation of MitraClip® Devices [EXPAND]; NCT03502811).

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jcin.2023.01.010DOI Listing

Publication Analysis

Top Keywords

transcatheter edge-to-edge
12
edge-to-edge repair
12
expand study
12
third-generation mitraclip
8
mitraclip ntr/xtr
8
ntr/xtr transcatheter
8
contemporary real-world
8
primary secondary
8
york heart
8
heart association
8

Similar Publications

Background: Up to 50% of patients undergoing mitral transcatheter edge-to-edge repair (MTEER) have an indication for left atrial appendage occlusion (LAAO). However, prospective evaluation of this strategy is lacking.

Objectives: The aim of the study was to prospectively evaluate the outcomes of combined LAAO and MTEER.

View Article and Find Full Text PDF

Objective: This meta-analysis aims to assess the impact of pulmonary hypertension (PH) on the clinical prognosis of patients with moderate to severe mitral valve regurgitation (MR) undergoing transcatheter edge-to-edge mitral valve repair (TEER).

Methods: As of August 2024, relevant studies were identified through searches of the PubMed, Cochrane Library, Web of Science, and Embase databases. A comprehensive screening process was conducted, with quality assessment performed utilizing the Newcastle Ottawa Scale (NOS).

View Article and Find Full Text PDF

Background: There are few reports about the one-stage surgery of transcatheter aortic valve replacement (TAVR) + mitral valve transcatheter edge-to-edge repair (M-TEER) around the world. TAVR + M-TEER surgery is usually performed under the simultaneous guidance of digital subtraction angiography (DSA) and echocardiography. There is no report of TAVR surgery assisted only by echocardiography all over the world.

View Article and Find Full Text PDF

Temporal profiling of M-TEER-related complications.

ESC Heart Fail

January 2025

Division of Cardiology, Pulmonology and Vascular Medicine, Medical Faculty, University of Düsseldorf, Düsseldorf, Germany.

Aims: Transcatheter edge-to-edge repair of the mitral valve (M-TEER) is known for its low complication rates. However, the optimal level and duration of post-procedural care remain unclear. This study aimed to identify the specific timeframe of post-procedural complications following M-TEER.