Background: Traditional reconstruction techniques have certain limitations in balancing image quality and reducing radiation dose. The deep learning image reconstruction (DLIR) algorithm opens the door to a new era of medical image reconstruction. The purpose of the study was to evaluate the DLIR images at 1.25 mm thickness in balancing image noise and spatial resolution in low-dose abdominal computed tomography (CT) in comparison with the conventional adaptive statistical iterative reconstruction-V at 40% strength (ASIR-V40%) at 5 and 1.25 mm.

Methods: This retrospective study included 89 patients who underwent low-dose abdominal CT. Five sets of images were generated using ASIR-V40% at a 5 mm slice thickness and 1.25 mm (high-resolution) with DLIR at 1.25 mm using 3 strengths: low (DLIR-L), medium (DLIR-M), and high (DLIR-H). Qualitative evaluation was performed for image noise, artifacts, and visualization of small structures, while quantitative evaluation was performed for standard deviation (SD), signal-to-noise ratio (SNR), and spatial resolution (defined as the edge rising slope).

Results: At 1.25 mm, DLIR-M and DLIR-H images had significantly lower noise (SD in fat: 14.29±3.37 and 9.65±3.44 HU, respectively), higher SNR for liver (3.70±0.78 and 5.64±1.20, respectively), and higher overall image quality (4.30±0.44 and 4.67±0.40, respectively) than did the respective values in ASIR-V40% images (20.60±4.04 HU, 2.60±0.63, and 3.77±0.43; all P values <0.05). Compared with the 5 mm ASIR-V40% images, the 1.25 mm DLIR-H images had lower noise (SD: 9.65±3.44 13.63±10.03 HU), higher SNR (5.64±1.20 4.69±1.28), and higher overall image quality scores (4.67±0.40 3.94±0.46) (all P values <0.001). In addition, DLIR-L, DLIR-M, and DLIR-H images had a significantly higher spatial resolution in terms of edge rising slope (59.66±21.46, 58.52±17.48, and 59.26±13.33, respectively, 33.79±9.23) and significantly higher image quality scores in the visualization of fine structures (4.43±0.50, 4.41±0.49, and 4.38±0.49, respectively 2.62±0.49) than did the 5 mm ASIR-V40 images.

Conclusions: The 1.25 mm DLIR-M and DLIR-H images had significantly reduced image noise and improved SNR and overall image quality compared to the 1.25 mm ASIR-V40% images, and they had significantly improved the spatial resolution and visualization of fine structures compared to the 5 mm ASIR-V40% images. DLIR-H images had further reduced image noise compared with the 5 mm ASIR-V40% images, and DLIR-H was the most effective technique at balancing the image noise and spatial resolution in low-dose abdominal CT.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10006151PMC
http://dx.doi.org/10.21037/qims-22-353DOI Listing

Publication Analysis

Top Keywords

image reconstruction
12
image noise
12
spatial resolution
12
low-dose abdominal
12
deep learning
8
image
8
learning image
8
slice thickness
8
noise spatial
8
resolution low-dose
8

Similar Publications

Objectives: To determine whether deep learning-based reconstructions of zero-echo-time (ZTE-DL) sequences enhance image quality and bone visualization in cervical spine MRI compared to traditional zero-echo-time (ZTE) techniques, and to assess the added value of ZTE-DL sequences alongside standard cervical spine MRI for comprehensive pathology evaluation.

Methods: In this retrospective study, 52 patients underwent cervical spine MRI using ZTE, ZTE-DL, and T2-weighted 3D sequences on a 1.5-Tesla scanner.

View Article and Find Full Text PDF

Purpose: Given the potential role of nasolacrimal duct (NLD) morphometry in the aetiology of primary acquired obstructions, it is imperative that clinicians have access to detailed anatomical information. The aim of this study was to determine normative data on nasolacrimal duct morphometry in the Turkish population sample and to provide guidelines for clinicians.

Methods: The study included retrospectively computed tomography images of a healthy Turkish population sample of 151 individuals, 79 of whom were female and 72 of whom were male.

View Article and Find Full Text PDF

Background: To define optimal parameters for the evaluation of vessel visibility in intracranial stents (ICS) and flow diverters (FD) using photon-counting detector computed tomography angiography (PCD-CTA) with spectral reconstructions.

Methods: We retrospectively analyzed consecutive patients with implanted ICS or FD, who received a PCD-CTA between April 2023 and March 2024. Polyenergetic, virtual monoenergetic, pure lumen, and iodine reconstructions with different keV levels (40, 60, and 80) and reconstruction kernels (body vascular [Bv]48, Bv56, Bv64, Bv72, and Bv76) were evaluated by two radiologists with regions of interests and Likert scales.

View Article and Find Full Text PDF

Indian agriculture is vital sector in the country's economy, providing employment and sustenance to millions of farmers. However, Plant diseases are a serious risk to crop yields and farmers' livelihoods. Traditional plant disease diagnosis methods rely heavily on human expertise, which can lead to inaccuracies due to the invisible nature of early disease symptoms and the labor-intensive process, making them inefficient for large-scale agricultural management.

View Article and Find Full Text PDF

Optimizing dental implant identification using deep learning leveraging artificial data.

Sci Rep

January 2025

Department of Oral and Maxillofacial Surgery, Faculty of Medicine, Kagawa University, 1750-1, Ikenobe, Miki-cho, Kita-gun, Takamatsu, 761-0793, Kagawa, Japan.

This study aims to evaluate the potential enhancement in implant classification performance achieved by incorporating artificially generated images of commercially available products into a deep learning process of dental implant classification using panoramic X-ray images. To supplement an existing dataset of 7,946 in vivo dental implant images, a three-dimensional scanner was employed to create implant surface models. Subsequently, implant surface models were used to generate two-dimensional X-ray images, which were compiled along with original images to create a comprehensive dataset.