Background: The solid-state cadmium-zinc-telluride (CZT) gamma camera for myocardial perfusion single-photon emission computed tomography (MPS) has theoretical advantages compared to the conventional gamma camera technique. This includes more sensitive detectors and better energy resolution. We aimed to explore the diagnostic performance of gated MPS with a CZT gamma camera compared to a conventional gamma camera for detection of myocardial infarct (MI) and assessment of left ventricular (LV) volumes and ejection fraction (LVEF), using cardiac magnetic resonance (CMR) as the reference method.

Methods: Seventy-three patients (26% female) with known or suspected chronic coronary syndrome were examined with gated MPS using both a CZT gamma camera and a conventional gamma camera as well as with CMR. Presence and extent of MI on MPS and late gadolinium enhancement (LGE) CMR was evaluated. For LV volumes, LVEF and LV mass, gated MPS images and cine CMR images were evaluated.

Results: MI was found in 42 patients on CMR. The overall sensitivity, specificity, positive and negative predictive values for the CZT and the conventional gamma camera were the same (67%, 100%, 100% and 69%). For infarct size > 3% on CMR, the sensitivity was 82% for the CZT and 73% for the conventional gamma camera, respectively. LV volumes were significantly underestimated by MPS compared to CMR (P ≤ .002 for all measures). The underestimation was slightly less pronounced for the CZT compared to the conventional gamma camera (2-10 mL, P ≤ .03 for all measures). For LVEF, however, accuracy was high for both gamma cameras.

Conclusion: Differences between a CZT and a conventional gamma camera for detection of MI and assessment of LV volumes and LVEF are small and do not appear to be clinically significant.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10558368PMC
http://dx.doi.org/10.1007/s12350-022-03185-0DOI Listing

Publication Analysis

Top Keywords

gamma camera
48
conventional gamma
32
czt gamma
16
compared conventional
16
gamma
13
camera
12
gated mps
12
czt
8
camera compared
8
conventional
8

Similar Publications

GFR Estimation and Correlation for Oncology Patients by Two Methods, Gates Method and Dual Time Point Plasma Sampling Method.

Indian J Nucl Med

November 2024

Department of Nuclear Medicine and Molecular Imaging, Homi Bhabha Cancer Hospital and Mahamana Pandit Madan Mohan Malaviya Cancer Centre, Tata Memorial Centre, Homi Bhabha National Institute (HBNI), Varanasi, India.

Background: With the increasing number of oncology cases and a parallel surge in chemotherapeutic drugs for treatment, the treating physicians conducts nephrotoxicity evaluation to provide a personalized dosing strategy. Of the various tests available, glomerular filtration rate (GFR) under gamma camera with help of Gates method has gained importance, being a good index of overall kidney functions. In addition to this, there has been an alternate and old method for GFR estimation: plasma sampling.

View Article and Find Full Text PDF

Background: High-dose-rate (HDR) brachytherapy using Iridium-192 as a radiation source is widely employed in cancer treatment to deliver concentrated radiation doses while minimizing normal tissue exposure. In this treatment, the precision with which the sealed radioisotope source is delivered significantly impacts clinical outcomes.

Purpose: This study aims to evaluate the feasibility of a new four-dimensional (4D) in vivo source tracking and treatment verification system for HDR brachytherapy using a patient-specific approach.

View Article and Find Full Text PDF

cutaneous melanoma has often unpredictable lymphatic drainage patterns, especially at the level of the trunk, head and neck regions. Sentinel lymph node biopsy (SLNB) is an important prognostic tool that accurately assesses regional lymph node involvement and guides therapeutic decisions. Material and this prospective study involved 104 patients diagnosed with cutaneous melanoma who underwent SLNB using a radioactive tracer.

View Article and Find Full Text PDF

Autonomous vehicles, often known as self-driving cars, have emerged as a disruptive technology with the promise of safer, more efficient, and convenient transportation. The existing works provide achievable results but lack effective solutions, as accumulation on roads can obscure lane markings and traffic signs, making it difficult for the self-driving car to navigate safely. Heavy rain, snow, fog, or dust storms can severely limit the car's sensors' ability to detect obstacles, pedestrians, and other vehicles, which pose potential safety risks.

View Article and Find Full Text PDF

Introduction And Objectives: Tissue attenuation reduces the specificity of the myocardial perfusion imaging single photon emission tomography (SPECT), which leads reduced diagnostic accuracy. The aim of this study is to compare performances of non-attenuation corrected (NAC), computed tomography based-attenuation corrected (AC) and prone images for qualitative and semi-quantitative analysis of myocardial perfusion SPECT in diagnosis of coronary artery disease (CAD).

Materials And Methods: Eightysix patients in whom NAC, AC and prone images were obtained with SPECT at Selcuk University Faculty of Medicine, and whose coronary angiography/CT coronary angiography was completed within 3 months, were retrospectively studied.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!