Impact of the new ultra-high sensitivity mode in a long axial field-of-view PET/CT.

Ann Nucl Med

Department of Nuclear Medicine, Inselspital, Bern University Hospital, University of Bern, Street: Freiburgstr. 18, 3010, Bern, Switzerland.

Published: May 2023

Objective: Long axial field-of-view (LAFOV) PET/CT showed improved performance resulting from higher sensitivity. The aim was to quantify the impact of using the full acceptance angle (UHS) in image reconstructions with the Biograph Vision Quadra LAFOV PET/CT (Siemens Healthineers) compared to the limited acceptance angle (high sensitivity mode, HS).

Methods: 38 oncological patients examined on a LAFOV Biograph Vision Quadra PET/CT were analysed. 15 patients underwent [F]FDG-PET/CT, 15 patients underwent [F]PSMA-1007 PET/CT, and 8 patients underwent [Ga]Ga-DOTA-TOC PET/CT. Signal-to-noise ratio (SNR) and standardised uptake values (SUV) were used to compare UHS and HS with different acquisition times.

Results: The SNR was significantly higher for UHS compared to HS over all acquisition times (SNR UHS/HS [F]FDG: 1.35 ± 0.02, p < 0.001; [F]PSMA-1007: 1.25 ± 0.02, p < 0.001; [Ga]Ga-DOTA-TOC: 1.29 ± 0.02, p < 0.001).

Conclusion: UHS showed significantly higher SNR opening the possibility of halving short acquisition times. This is of advantage in further reduction of whole-body PET/CT acquisition.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10129991PMC
http://dx.doi.org/10.1007/s12149-023-01827-yDOI Listing

Publication Analysis

Top Keywords

patients underwent
12
sensitivity mode
8
long axial
8
axial field-of-view
8
lafov pet/ct
8
acceptance angle
8
biograph vision
8
vision quadra
8
pet/ct
6
impact ultra-high
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!