Background: Aortic arch pathology often requires staged segmental repairs. Total aortic arch replacement with frozen elephant trunk (FET) offers surgical options for these pathologies. The Jotec E-vita Open NEO™ branched prosthesis was introduced in 2020; we sought to share our initial experience focusing on the prosthesis selection strategies, surgical techniques, anastomosis-bleeding and graft-oozing control methods, and early clinical outcomes from two Asian centers.

Methods: We performed a retrospective cohort study in patients with aortic arch pathologies who underwent total arch replacement using the FET procedure with Jotec E-vita Open NEO™ branched prosthesis from two Asian centers between October 2020 and August 2021. The primary outcome was overall 30-day mortality, and the secondary outcomes were operative complications.

Results: Twenty-five consecutive patients underwent total arch replacement with FET with the novel hybrid prosthesis. Overall 30-day mortality from both centers was 0%. Overall mean operative, cardiopulmonary bypass, hypothermic circulatory arrest, and selective antegrade cerebral perfusion times were 353.4±80.5, 183.2±39.6, 57.2±14.7, and 138.2±28.6 minutes, respectively. No patient developed stroke. Permanent spinal cord injury (SCI) was recorded in one patient (4%) and one (4%) had transient lower limb weakness that resolved after spinal drainage. There was no requirement of re-sternotomy for hemostasis.

Conclusions: We reported a multicenter Asian case series with the novel FET hybrid prosthesis demonstrating the feasibility and safety of promising initial clinical outcomes. The technique of circumferential reinforcement of vascular anastomosis for hemostasis may be one of the methods for lowering the rates of re-sternotomy for hemostasis, and proper surgical or transfusion strategies would overcome the excessive oozing of the prosthesis. Long-term follow-up is required for further evaluation of aortic pathology progression and device-related outcomes.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9992573PMC
http://dx.doi.org/10.21037/jtd-22-1055DOI Listing

Publication Analysis

Top Keywords

e-vita open
12
aortic arch
12
arch replacement
12
jotec e-vita
8
open neo™
8
neo™ branched
8
branched prosthesis
8
clinical outcomes
8
underwent total
8
total arch
8

Similar Publications

Article Synopsis
  • The Ascyrus Medical Dissection Stent (AMDS) can complicate the management of acute type A aortic dissection (ATAAD) due to issues with the false lumen (FL) persisting after treatment.
  • Two case studies illustrated the complications from failed AMDS treatments, leading to the successful use of a trifurcated hybrid prosthesis and frozen elephant trunk (FET) technique for redo aortic arch repair.
  • This new approach has shown potential in improving patient outcomes by effectively addressing complications from failed treatments and offering better long-term survival rates.
View Article and Find Full Text PDF

Objectives: The aim of this multicentre study was to demonstrate the safety and clinical performance of E-vita OPEN NEO Stent Graft System (Artivion, Inc.) in the treatment of aneurysm or dissection, both acute and chronic, in the ascending aorta, aortic arch and descending thoracic aorta.

Methods: In this observational study of 12 centres performed in Europe and in Asia patients were enrolled between December 2020 and March 2022.

View Article and Find Full Text PDF

Surgery for acute type A aortic dissection is highly challenging, even in expert hands. The goal in such emergency circumstances is primarily to save the patient's life. To minimize the perioperative risk, surgeons often choose surgery involving only supracoronary ascending aortic and hemiarch replacement.

View Article and Find Full Text PDF

Objectives: In this cohort study, we aimed to assess the 1-year clinical outcomes of using the E-vita Open NEO™ hybrid prosthesis for total arch replacement with frozen elephant trunk (FET) to repair extensive aortic pathologies.

Methods: We reviewed individuals who underwent thoracic aortic surgery between April 2021 and March 2023 from the Gangnam Severance Aortic Registry. Exclusion criteria included ascending aortic replacement, 1 or 2 partial arch replacement, descending aortic replacement and total arch replacement without an FET.

View Article and Find Full Text PDF

Introduction: Recent reports have questioned the blood impermeability of the novel frozen elephant trunk (FET) device E-vita Open NEO (EO-NEO). Therefore, standardized bleeding tests using porcine heparinized blood were performed, as well as stress testing on the blood tightness of the collar suture line, to investigate this observation.

Material And Methods: EO-NEO prostheses were examined for blood permeability in three test series.

View Article and Find Full Text PDF