Background: The geometry and compliance of gastrointestinal sphincters may be assessed by impedance planimetry using a functional lumen imaging probe (FLIP). We describe our institutional foregut surgeon experience using FLIP in 1,097 cases, highlighting instances where FLIP changed operative decision making.
Study Design: A retrospective review of an IRB-approved prospective quality database was performed. This included operative and endoscopic suite foregut procedures using FLIP between February 2013 and May 2022.
Results: During the study period, FLIP was used a total of 1,097 times in 919 unique patients by 2 foregut surgeons. Intraoperative FLIP was used during 573 antireflux procedures and 272 endoscopic myotomies. FLIP was also used during 252 endoscopic suite procedures. For those undergoing preoperative workup of GERD, starting in 2021, esophageal panometry was performed in addition to standard FLIP measurements at the lower esophageal sphincter. In 77 cases, intraoperative FLIP changed operative decision making. During antireflux procedures, changes included adding or removing crural sutures, adjusting a fundoplication tightness, choice of full vs partial wrap, and magnetic sphincter augmentation sizing. For endoscopic procedures, changes included aborting peroral endoscopic myotomy or Zenker's peroral endoscopic myotomy, performing a myotomy when preoperative diagnosis was unclear, or performing additional myotomy.
Conclusions: FLIP is a useful tool for assessing the upper esophageal sphincter, lower esophageal sphincter, pylorus, and secondary esophageal peristalsis that can be used in a wide variety of clinical situations within a foregut surgeon's practice. It can also function as an adjunct in intraoperative decision making.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/XCS.0000000000000675 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!