Background: Initial data about the performance of the new-generation SAPIEN 3 Ultra (S3U) valve are highly promising. However, evidence about the longer-term performance and safety of the S3U is scarce.

Aims: We aimed to investigate the 1-year clinical and echocardiographic outcomes of transcatheter aortic valve implantation (TAVI) using the S3U compared with its predecessor, the SAPIEN 3 valve (S3).

Methods: The SAPIEN 3 Ultra registry included consecutive patients who underwent transfemoral TAVI at 12 European centres with the S3U or S3 between October 2016 and December 2020. One-to-one propensity score (PS) matching was performed to account for differences in baseline characteristics. The primary outcomes of interest were all-cause death and the composite of all-cause death, disabling stroke and hospitalisation for heart failure at 1 year.

Results: The overall study cohort encompassed 1,692 patients treated with either the S3U (n=519) or S3 (n=1,173). The PS-matched population had a total of 992 patients (496 per group). At 1 year, the rate of death from any cause was 4.9% in the S3U group and 6.3% in the S3 group (p=0.743). Similarly, there were no significant differences in the rates of the primary composite outcome (9.5% in the S3 group and 6.6% in the S3U group; p=0.162). The S3U was associated with lower rates of mild paravalvular leak (PVL) compared with the S3 (odds ratio 0.63, 95% confidence interval: 0.44 to 0.88; p<0.01). No significant differences in transprosthetic gradients were observed between the two groups.

Conclusions: Compared with the S3, the S3U transcatheter heart valve was associated with similar 1-year clinical outcomes but reduced rates of mild PVL.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10111124PMC
http://dx.doi.org/10.4244/EIJ-D-22-01022DOI Listing

Publication Analysis

Top Keywords

s3u
9
outcomes transcatheter
8
transcatheter aortic
8
aortic valve
8
valve implantation
8
sapien ultra
8
all-cause death
8
s3u group
8
valve
5
group
5

Similar Publications

Background: Longer term outcomes of transcatheter aortic valve replacement (TAVR) with the SAPIEN 3 Ultra RESILIA (S3UR) valve over its predecessors have not been reported in a national registry.

Objectives: The aim of this study was to compare the 1-year clinical and echocardiographic outcomes of the S3UR with those of the SAPIEN 3 (S3) and SAPIEN 3 Ultra (S3U) in the Society of Thoracic Surgeons/American College of Cardiology TVT (Transcatheter Valve Therapy) Registry.

Methods: From September 2022 to March 2023, patients who underwent native TAVR with the S3UR or the S3 or S3U valve were propensity matched and compared.

View Article and Find Full Text PDF

Background: The fifth-generation SAPIEN 3 Ultra Resilia valve (S3UR) incorporates several design changes as compared with its predecessors, the SAPIEN 3 (S3) and SAPIEN 3 Ultra (S3U) valves, including bovine leaflets treated with a novel process intended to reduce structural valve deterioration via calcification, as well as a taller external skirt on the 29-mm valve size to reduce paravalvular leak (PVL). The clinical performance of S3UR compared with S3 and S3U in a large patient population has not been previously reported.

Objectives: The aim of this study was to compare S3UR to S3/S3U for procedural, in-hospital, and 30-day clinical and echocardiographic outcomes after transcatheter aortic valve replacement (TAVR).

View Article and Find Full Text PDF
Article Synopsis
  • The study investigates the optimal percent oversizing (%OS) for the SAPIEN3 Ultra (S3U) transcatheter aortic valve to minimize paravalvular regurgitation (PVR) while assessing the risk of conduction disturbances (CD).
  • It compared 350 patients with the S3U against 606 patients using the original SAPIEN3 (S3), finding significantly lower rates of PVR in the S3U group across various %OS levels.
  • The findings suggest that a minimal %OS (0% to 5%) is optimal for the S3U, balancing the risks of PVR and CD more effectively than the S3.
View Article and Find Full Text PDF

Latest-iteration balloon- and self-expandable transcatheter valves for severe bicuspid aortic stenosis: the TRITON study.

Rev Esp Cardiol (Engl Ed)

November 2023

Departamento de Cardiología, Hospital Clínico Universitario de Valladolid, Valladolid, Spain; Centro de Investigación Biomédica en Red de Enfermedades Cardiovasculares, Spain.

Introduction And Objectives: No comparisons have been published yet regarding the newest iteration of balloon- and self-expandable transcatheter heart valves for the treatment of bicuspid aortic valve (BAV) stenosis.

Methods: Multicenter registry of consecutive patients with severe BAV stenosis treated with balloon-expandable transcatheter heart valves (Myval and SAPIEN 3 Ultra, S3U) or self-expanding Evolut PRO+(EP+). TriMatch analysis was carried out to minimize the impact of baseline differences.

View Article and Find Full Text PDF

Background: Initial data about the performance of the new-generation SAPIEN 3 Ultra (S3U) valve are highly promising. However, evidence about the longer-term performance and safety of the S3U is scarce.

Aims: We aimed to investigate the 1-year clinical and echocardiographic outcomes of transcatheter aortic valve implantation (TAVI) using the S3U compared with its predecessor, the SAPIEN 3 valve (S3).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!