Objective: To determine the safety and effectiveness of vena cava filters (VCFs).
Methods: A total of 1429 participants (62.7 ± 14.7 years old; 762 [53.3% male]) consented to enroll in this prospective, nonrandomized study at 54 sites in the United States between October 10, 2015, and March 31, 2019. They were evaluated at baseline and at 3, 6, 12, 18, and 24 months following VCF implantation. Participants whose VCFs were removed were followed for 1 month after retrieval. Follow-up was performed at 3, 12, and 24 months. Predetermined composite primary safety (freedom from perioperative serious adverse events [AEs] and from clinically significant perforation, VCF embolization, caval thrombotic occlusion, and/or new deep vein thrombosis [DVT] within 12-months) and effectiveness (composite comprising procedural and technical success and freedom from new symptomatic pulmonary embolism [PE] confirmed by imaging at 12-months in situ or 1 month postretrieval) end points were assessed.
Results: VCFs were implanted in 1421 patients. Of these, 1019 (71.7%) had current DVT and/or PE. Anticoagulation therapy was contraindicated or had failed in 1159 (81.6%). One hundred twenty-six (8.9%) VCFs were prophylactic. Mean and median follow-up for the entire population and for those whose VCFs were not removed was 243.5 ± 243.3 days and 138 days and 332.6 ± 290 days and 235 days, respectively. VCFs were removed from 632 (44.5%) patients at a mean of 101.5 ± 72.2 days and median 86.3 days following implantation. The primary safety end point and primary effectiveness end point were both achieved. Procedural AEs were uncommon and usually minor, but one patient died during attempted VCF removal. Excluding strut perforation greater than 5 mm, which was demonstrated on 31 of 201 (15.4%) patients' computed tomography scans available to the core laboratory, and of which only 3 (0.2%) were deemed clinically significant by the site investigators, VCF-related AEs were rare (7 of 1421, 0.5%). Postfilter, venous thromboembolic events (none fatal) occurred in 93 patients (6.5%), including DVT (80 events in 74 patients [5.2%]), PE (23 events in 23 patients [1.6%]), and/or caval thrombotic occlusions (15 events in 15 patients [1.1%]). No PE occurred in patients following prophylactic placement.
Conclusions: Implantation of VCFs in patients with venous thromboembolism was associated with few AEs and with a low incidence of clinically significant PEs.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jvsv.2022.11.002 | DOI Listing |
Neurospine
December 2024
Department of Neurosurgery, Gachon University Gil Medical Center, Incheon, Korea.
We present a case of iatrogenic lumbar foraminal stenosis caused by bone-cement leakage during vertebroplasty, successfully managed using transforaminal endoscopic lumbar foraminotomy (TELF). Vertebroplasty is an effective treatment for osteoporotic vertebral compression fractures (VCFs); however, complications such as bone-cement leakage can lead to vascular or neurological issues, including lumbar radiculopathy. TELF is a minimally invasive surgical option for addressing various forms of lumbar foraminal stenosis.
View Article and Find Full Text PDFCell Death Differ
December 2024
Georgetown University Medical Center, Lombardi Comprehensive Cancer Center, Washington, D.C., USA.
Germline inactivating mutations of the SLC25A1 gene contribute to various human disorders, including Velocardiofacial (VCFS), DiGeorge (DGS) syndromes and combined D/L-2-hydroxyglutaric aciduria (D/L-2HGA), a severe systemic disease characterized by the accumulation of 2-hydroxyglutaric acid (2HG). The mechanisms by which SLC25A1 loss leads to these syndromes remain largely unclear. Here, we describe a mouse model of SLC25A1 deficiency that mimics human VCFS/DGS and D/L-2HGA.
View Article and Find Full Text PDFJ Vasc Surg Venous Lymphat Disord
January 2024
Division of Interventional Cardiology, Hospital Alemán, Buenos Aires, Argentina; Division of Interventional Cardiology, Hospital Británico de Buenos Aires, Buenos Aires, Argentina.
Objective: Inferior vena cava filters (VCFs) are a therapeutic resource for the treatment of patients with thromboembolic disease who have a contraindication to full-dose anticoagulation. In the present study, we report the retrieval rate and long-term mortality of patients receiving optional inferior VCFs and identify the predictors for retrieval and all-cause mortality during follow-up.
Methods: We conducted a retrospective cohort study of 739 consecutive recipients of optional inferior VCFs from January 2002 to December 2021 in two hospitals.
J Vasc Surg Venous Lymphat Disord
May 2023
Department of Vascular and Endovascular Surgery, Beth Israel Deaconess Medical Center, Brockton, MA.
Objective: To determine the safety and effectiveness of vena cava filters (VCFs).
Methods: A total of 1429 participants (62.7 ± 14.
J Vasc Interv Radiol
April 2023
Department of Vascular and Endovascular Surgery, Beth Israel Deaconess Medical Center, Brockton, MA.
Objective: To determine the safety and effectiveness of vena cava filters (VCFs).
Methods: A total of 1429 participants (62.7 ± 14.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!