High-resolution 3D ultra-short echo time MRI with Rosette k-space pattern for brain iron content mapping.

J Trace Elem Med Biol

Weldon School of Biomedical Engineering, Purdue University, West Lafayette, IN, USA; School of Health Sciences, Purdue University, West Lafayette, IN, USA. Electronic address:

Published: May 2023

Background: The iron concentration increases during normal brain development and is identified as a risk factor for many neurodegenerative diseases, it is vital to monitor iron content in the brain non-invasively.

Purpose: This study aimed to quantify in vivo brain iron concentration with a 3D rosette-based ultra-short echo time (UTE) magnetic resonance imaging (MRI) sequence.

Methods: A cylindrical phantom containing nine vials of different iron concentrations (iron (II) chloride) from 0.5 millimoles to 50 millimoles and six healthy subjects were scanned using 3D high-resolution (0.94 ×0.94 ×0.94 mm) rosette UTE sequence at an echo time (TE) of 20 μs.

Results: Iron-related hyperintense signals (i.e., positive contrast) were detected based on the phantom scan, and were used to establish an association between iron concentration and signal intensity. The signal intensities from in vivo scans were then converted to iron concentrations based on the association. The deep brain structures, such as the substantia nigra, putamen, and globus pallidus, were highlighted after the conversion, which indicated potential iron accumulations.

Conclusion: This study suggested that T-weighted signal intensity could be used for brain iron mapping.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10107748PMC
http://dx.doi.org/10.1016/j.jtemb.2023.127146DOI Listing

Publication Analysis

Top Keywords

echo time
12
brain iron
12
iron concentration
12
iron
10
ultra-short echo
8
iron content
8
iron concentrations
8
signal intensity
8
brain
6
high-resolution ultra-short
4

Similar Publications

Background: The pressure gradient between the ventricles and the subarachnoid space (transmantle pressure) is crucial for understanding CSF circulation and the pathogenesis of certain neurodegenerative diseases. This pressure can be approximated by the pressure difference across the aqueduct (ΔP). Currently, no dedicated platform exists for quantifying ΔP, and no research has been conducted on the impact of breathing on ΔP.

View Article and Find Full Text PDF

Objectives: To explore the reproducibility of the 1.5-T MR imaging (MRI)-based R2* method in measuring the liver iron concentration (LIC) across different MRI scanners, scan parameters, and postprocessing techniques.

Materials And Methods: We performed a systematic search of the PubMed, Embase, Medline, Cochrane Library, and Web of Science databases and identified studies that used the 1.

View Article and Find Full Text PDF

Background: Brain arteriolosclerosis is characterized by the thickening of vessel walls and arteriolar stenosis and is one of the primary pathologies of cerebral small vessel disease. Arteriolosclerosis is linked to lower cognitive and motor function, as well as an elevated risk of dementia. This study aimed to investigate the association of brain arteriolosclerosis with regional gray matter volumes in a large number of community-based older adults.

View Article and Find Full Text PDF

Background: We evaluated the accuracy of magnetic resonance imaging (MRI) computed tomography (CT)-like sequences compared to normal-resolution CT (NR-CT) and super-high-resolution CT (SHR-CT) for planning of cochlear implantation.

Methods: Six cadaveric temporal bone specimens were used. 3-T MRI scans were performed using radial volumetric interpolated breath-hold (STARVIBE), pointwise-encoding time reduction with radial acquisition (PETRA), and ultrashort time of echo (UTE) sequences.

View Article and Find Full Text PDF

Artifacts at Cardiac MRI: Imaging Appearances and Solutions.

Radiographics

January 2025

From the Department of Radiology, Cardiovascular Imaging, Mayo Clinic, 200 1st St SW, Rochester, MN 559905 (P.S.R., P.A.A.); Department of Radiology, Division of Cardiothoracic Imaging, Jefferson University Hospitals, Philadelphia, Pa (B.S.); Department of Radiology, Baylor Health System, Dallas, Tex (P.R.); Department of Diagnostic Radiology, School of Clinical Medicine, Li Ka Shing Faculty of Medicine, University of Hong Kong, Hong Kong SAR (M.Y.N.); and Department of Diagnostic Radiology, Cleveland Clinic, Cleveland, Ohio (M.A.B.).

Cardiac MRI (CMR) is an important imaging modality in the evaluation of cardiovascular diseases. CMR image acquisition is technically challenging, which in some circumstances is associated with artifacts, both general as well as sequence specific. Recognizing imaging artifacts, understanding their causes, and applying effective approaches for artifact mitigation are critical for successful CMR.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!