Agreement of Oscillometric and Auscultatory blood pressure measurement methods: An ambulance noise simulation study.

Am J Emerg Med

University of Health Sciences, Dept. of Emergency Medicine, Kartal Dr. Lutfi Kirdar City Hospital, Istanbul, Turkey. Electronic address:

Published: May 2023

Objective: Although noise is known to negatively affect blood pressure (BP) measurements, its impact on different BP measurement methods remains unclear. The aim of this study is to compare the agreement of oscillometric and auscultatory BP measurement methods under in-ambulance noise levels.

Methods: This method-comparison study was conducted on 50 healthy volunteers in a tertiary emergency department (ED). Participants were divided into two groups of 25, and BP was measured using auscultatory and oscillometric methods in noisy and ambient environments by 2 emergency medicine technicians (EMT). The primary object of the study was to compare the agreement of auscultatory mercury sphygmomanometers and automated auscillometric BP measurements in ambient and noisy environments.

Results: We examined the agreement between auscultative and oscillometric measurements of BP conducted in an ambient environment (46.75 [IQR (41.2--55.18)] dB) and found that both systolic and diastolic BP were within the level of agreement (LoA) established before the study (systolic BP [-13.96 to 8.48 mmHG], diastolic BP [-7.44 to 8.08 mmHg]); whereas, in noisy environment (92.35 [IQR 88-96.55] dB) both systolic and diastolic BP were outside the range of LoA (systolic BP [-37.77 to 9.94 mmHg], diastolic BP [-21.73 to 16.37 mmHg]). Additionally, we found that in ambient environments, concordance correlation coefficients were higher than in noisy environments (0.943 [0.906-0.966], 0.957 [0.93-0.974]; 0.574 [0.419-0.697], 0.544 [0.326-0.707]; systolic and diastolic BP, respectively).

Conclusion: The results of this study demonstrate that noise significantly affects the agreement between oscillometric and auscultatory blood pressure measurement methods.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ajem.2023.02.022DOI Listing

Publication Analysis

Top Keywords

measurement methods
16
agreement oscillometric
12
oscillometric auscultatory
12
blood pressure
12
systolic diastolic
12
auscultatory blood
8
pressure measurement
8
study compare
8
compare agreement
8
ambient environments
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!