How a decade of aneurysms embolization with the Woven EndoBridge has changed our understanding and practices?

J Neuroradiol

NEURI Brain Vascular Center, Interventional Neuroradiology, Bicêtre University-Hospital, Le Kremlin-Bicêtre, France; Paris-Saclay University Faculty of Medicine, INSERM U1195, Le Kremlin-Bicêtre, France.

Published: September 2023

Background: Safety and efficacy of the WEB (Woven EndoBridge) device have been extensively evaluated in several good clinical practice studies. Nonetheless, the WEB had several structural evolutions overtime up to the fifth generation WEB device (WEB17). Here, we tried to understand how this may have modified our practices and enlarged our indications for its use.

Methods: We retrospectively analyzed data from all patients with aneurysms treated (or intended to be treated) with a WEB at our institution between July 2012 and February 2022. The time frame was split in two periods: before and after the arrival of the WEB17 in our center (February 2017).

Results: 252 patients with 276 wide-necked aneurysms were included; 78 aneurysms (28.2%) were ruptured. Successful embolization with a WEB device was achieved in 263/276 aneurysms (95.3%). With the availability of WEB17, treated aneurysms were significantly smaller (8.2 mm versus 5.9 mm, p<0.001) and off-label location significantly increased (4.4% versus 17.3%, p = 0.02) with an increase of sidewall aneurysm (4.4% versus 11.6%, p = 0.06). Also, WEB were more significantly oversized (1.05 versus 1.11, p<0.01). Adequate and complete occlusion rates increased constantly during the two periods (54.8% versus 67.5%, p = 0.08 and 74.2% versus 83.7%, p = 0.10, respectively). The proportion of ruptured aneurysms slightly increased between the two periods (24.6% versus 29.5%, p = 0.44).

Conclusions: Over the first decade of its availability, the WEB device usage shifted towards smaller aneurysms and broader indications, including ruptured aneurysms. The oversizing strategy also became the standard of practice for WEB deployment in our institution.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.neurad.2023.02.006DOI Listing

Publication Analysis

Top Keywords

woven endobridge
8
web device
8
web
5
aneurysms
5
decade aneurysms
4
aneurysms embolization
4
embolization woven
4
endobridge changed
4
changed understanding
4
understanding practices?
4

Similar Publications

The Woven EndoBridge (WEB) device has become a prominent treatment for wide-neck bifurcation intracranial aneurysms since its FDA approval in 2018. However, the impact of anticoagulant therapy on its efficacy and patient outcomes remains underexplored. This study aims to evaluate the effects of postoperative anticoagulant use on aneurysm occlusion, retreatment rates, and functional outcomes following WEB device implantation.

View Article and Find Full Text PDF

Purpose: Various angiographic assessment scales have been used to report the results of endovascular treatment with the WEB device. We aimed to review the use and reliability of these scales.

Methods: We systematically reviewed studies reporting angiographic outcomes of WEB-treated aneurysms from January 2010 to May 2023.

View Article and Find Full Text PDF

Background And Objectives: The Woven EndoBridge (WEB) device is a breakthrough in brain aneurysm treatment, widely used for intracranial wide neck and complex aneurysms, especially at bifurcations. Although effective, aneurysm recurrence can occur post-treatment and the impact of clipping angle on the WEB mesh remains unclear. In this study, using a human placenta-based ex-vivo aneurysm model, we simulated the application of aneurysm clipping on WEB-treated aneurysms to elucidate the technical aspects and nuances, particularly the influence of the angle of clip application on WEB collapse.

View Article and Find Full Text PDF

Purpose: The safety and efficacy of Woven EndoBridge (WEB) treatment has been proven. However, only a few standard techniques for safe and versatile WEB deployment have been described in the literature. In this study, we introduce the "make a flower bud and push at neck" technique to achieve safety and versatility during WEB treatment, referred to simply as the "flower bud" technique.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!