Improved lateral penumbra for proton ocular treatments on a general-purpose spot scanning beamline.

Phys Med

Department of Radiation Oncology, University of Washington School of Medicine, 1959 NE Pacific St., Seattle, WA 98195, USA; Fred Hutchinson Cancer Center, 1570 N 115th St., Seattle, WA 98133, USA.

Published: March 2023

Purpose: An ocular applicator that fits a commercial proton snout with an upstream range shifter to allow for treatments with sharp lateral penumbra is described.

Materials And Methods: The validation of the ocular applicator consisted of a comparison of range, depth doses (Bragg peaks and spread out Bragg peaks), point doses, and 2-D lateral profiles. Measurements were made for three field sizes, 1.5, 2, and 3 cm, resulting in 15 beams. Distal and lateral penumbras were simulated in the treatment planning system for seven range-modulation combinations for beams typical of ocular treatments and a field size of 1.5 cm, and penumbra values were compared to published literature.

Results: All the range errors were within 0.5 mm. The maximum averaged local dose differences for Bragg peaks and SOBPs were 2.6% and 1.1%, respectively. All the 30 measured point doses were within +/-3% of the calculated. The measured lateral profiles, analyzed through gamma index analysis and compared to the simulated, had pass rates greater than 96% for all the planes. The lateral penumbra increased linearly with depth, from 1.4 mm at 1 cm depth to 2.5 mm at 4 cm depth. The distal penumbra ranged from 3.6 to 4.4 mm and increased linearly with the range. The treatment time for a single 10 Gy (RBE) fractional dose ranged from 30 to 120 s, depending on the shape and size of the target.

Conclusions: The ocular applicator's modified design allows lateral penumbra similar to dedicated ocular beamlines while enabling planners to use modern treatment tools such as Monte Carlo and full CT-based planning with increased flexibility in beam placement.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ejmp.2023.102551DOI Listing

Publication Analysis

Top Keywords

lateral penumbra
16
bragg peaks
12
ocular treatments
8
ocular applicator
8
point doses
8
lateral profiles
8
increased linearly
8
penumbra
6
ocular
6
lateral
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!