Background: Although traditional surgical procedures for glaucoma (such as trabeculectomy and tube-shunt implantation) can significantly reduce intraocular pressure (IOP), they are associated with numerous complications, some of which are vision-threatening, or involve prolonged recovery or a highly intensive postoperative course. Micro-invasive glaucoma surgery (MIGS) procedures have shown better safety but reduced efficacy in achieving target IOP. Combinations of these methods have led to the development of subconjunctival micro-invasive procedures with safety comparable to traditional surgery and greater efficacy than minimally invasive methods. This review describes the use of one of these devices, the poly(styrene-block-isobutylene-block-styrene) (SIBS)-based PreserFlo MicroShunt (Santen, Emeryville, CA), in the surgical treatment of patients with glaucoma.

Main Text: The MicroShunt is an 8.5-mm tube made of an inert polymer with no endplate, an internal diameter of 70 μm, and fins intended to prevent peritubular flow and anchor the device within the sclera to prevent proximal migration into the eye. Following ab externo implantation, the tube provides a conduit for flow of aqueous humor from the anterior chamber into the subconjunctival/sub-Tenon space. Clinical trials to date have shown that, when paired with mitomycin C (MMC) treatment, MicroShunt implantation significantly reduced both IOP and the number of glaucoma medications. These IOP-lowering results were found both when surgery was performed alone and with phacoemulsification. The MicroShunt also showed a safety profile comparable to that of traditional filtering surgery.

Conclusions: The MicroShunt and other novel subconjunctival procedures have shown substantial IOP reductions while mitigating hypotony-related complications. MMC, which modulates fibrosis and scarring postoperatively, is essential to surgical success. Randomized, long-term clinical trials will further clarify the role of controlled micro-incisional device-assisted ab externo glaucoma filtering surgery in long-term glaucoma management.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9979478PMC
http://dx.doi.org/10.1186/s40662-022-00322-1DOI Listing

Publication Analysis

Top Keywords

microshunt implantation
8
comparable traditional
8
clinical trials
8
microshunt
6
glaucoma
6
subconjunctival filtration
4
filtration evolution
4
evolution current
4
current evidence
4
evidence microshunt
4

Similar Publications

Preserflo MicroShunt (PMS) implantation is a minimally invasive surgical procedure for treating glaucoma. Postoperative hypotony, a common complication of PMS implantation, can be prevented and treated with 10-0 nylon insertion. In this report, we present a case of postoperative hypotony following PMS implantation that was treated with intraluminal insertion of 9-0 nylon.

View Article and Find Full Text PDF

Background: Approximately 1.4 % of the German population aged 35 to 74 suffers from glaucoma, which is one of the more common causes of blindness. The only evidence-based treatment option at present is lowering the intraocular pressure.

View Article and Find Full Text PDF

Purpose: To investigate the efficacy and safety of the Preserflo MicroShunt (PMS) in treating glaucoma secondary to different uveitic glaucoma (UG) types in a European population.

Methods: This study evaluated consecutive eyes with UG that received the PMS. The primary outcome measure was 12-month surgical success (≥20% IOP reduction, target IOP of 6-21 mmHg).

View Article and Find Full Text PDF

Glaucoma and cataract often coexist. Patients with both conditions who qualify for surgical treatment may undergo either a combined surgical procedure or sequential treatments such as cataract surgery followed by an antiglaucoma procedure. A combined procedure with phacoemulsification is related to an increased risk of fibrosis of the filtering bleb; however, it is a rational approach for patients with high intraocular pressure and clinically significant lens opacification.

View Article and Find Full Text PDF

A 74-year-old Japanese woman with a history of primary open-angle glaucoma (POAG) in both eyes underwent PreserFlo MicroShunt (PFM) (Osaka, Japan: Santen Pharmaceutical, Co., Ltd.) implantation two years prior.

View Article and Find Full Text PDF