Background: Intensity-modulated helical tomotherapy (HT) is a promising technique in preparation for bone marrow transplantation. Nevertheless, radiation-sensitive organs can be substantially compromised due to suboptimal delivery techniques of total body irradiation (TBI). To reduce the potential burden of radiation toxicity to organs at risk (OAR), high-quality coverage and homogeneity are essential. We investigated dosimetric data from kidney, lung and thorax, liver, and spleen in relation to peripheral blood kinetics. To further advance intensity-modulated total body irradiation (TBI), the potential for dose reduction to lung and kidney was considered in the analysis.

Patients And Methods: 46 patients undergoing TBI were included in this analysis, partially divided into dose groups (2, 4, 8, and 12 Gy). HT was performed using a rotating gantry to ensuring optimal reduction of radiation to the lungs and kidneys and to provide optimal coverage of other OAR. Common dosimetric parameters, such as D05, D95, and D50, were calculated and analysed. Leukocytes, neutrophils, platelets, creatinine, GFR, haemoglobin, overall survival, and graft-versus-host disease were related to the dosimetric evaluation using statistical tests.

Results: The mean D95 of the lung is 48.23%, less than half the prescribed and unreduced dose. The D95 of the chest is almost twice as high at 84.95%. Overall liver coverage values ranged from 96.79% for D95 to 107% for D05. The average dose sparing of all patients analysed resulted in an average D95 of 68.64% in the right kidney and 69.31% in the left kidney. Average D95 in the spleen was 94.28% and D05 was 107.05%. Homogeneity indexes ranged from 1.12 for liver to 2.28 for lung. The additional significance analyses conducted on these blood kinetics showed a significant difference between the 2 Gray group and the other three groups for leukocyte counts. Further statistical comparisons of the dose groups showed no significant differences. However, there were significant changes in the dose of OAR prescribed with dose sparing (e.g., lung vs. rib and kidney).

Conclusion: Using intensity-modulated helical tomotherapy to deliver TBI is a feasible method in preparation for haematopoietic stem cell transplantation. Significant dose sparing in radiosensitive organs such as the lungs and kidneys is achievable with good overall quality of coverage. Peripheral blood kinetics support the positive impact of HT and its advantages strongly encourage its implementation within clinical routine.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10374741PMC
http://dx.doi.org/10.1007/s00432-023-04657-7DOI Listing

Publication Analysis

Top Keywords

blood kinetics
16
body irradiation
12
intensity-modulated helical
12
helical tomotherapy
12
dose sparing
12
dose
9
haematopoietic stem
8
stem cell
8
cell transplantation
8
organs risk
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!