Autonomous monitoring of orthophosphate (PO), an important nutrient for primary production in natural waters, is urgently needed. Here, we report on the development and validation of an autonomous electrochemical analyzer for PO in seawater. The approach is based on the use of flow injection analysis in conjunction with a dual electrochemical cell (i.e., a bi-potentiostat detector (FIA-DECD) that uses two working electrodes sharing the same reference and counter electrode. The two working electrodes are used (molybdate/carbon paste electrode (CPE) and CPE) to correct for matrix effects. Optimization of squarewave voltammetry parameters (including step potential, amplitude, and frequency) was undertaken to enhance analytical sensitivity. Possible interferences from non-ionic surfactants and humic acid were investigated. The limit of quantification in artificial seawater (30 g/L NaCl, pH 0.8) was 0.014 µM for a linear concentration range of 0.02-3 µM. The system used a Python script for operation and data processing. The analyzer was tested for ship-board PO determination during a four-day research cruise in the North Sea. The analyzer successfully measured 34 samples and achieved a good correlation (Pearson' R = 0.91) with discretely collected water samples analyzed using a laboratory-based colorimetric reference analyzer.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9959667 | PMC |
http://dx.doi.org/10.3390/s23042123 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!