The accurate detection of anti-neutralizing SARS-CoV-2 antibodies can aid in the understanding of the development of protective immunity against COVID-19. This study evaluated the diagnostic performance of the RapiSure (EDGC) COVID-19 S1 RBD IgG/Neutralizing Ab Test. Using the 90% plaque reduction neutralization test (PRNT) as a reference, 200 serum samples collected from 78 COVID-19-positive and 122 COVID-19-negative patients were divided into 76 PRNT-positive and 124 PRNT-negative groups. The ability of the RapiSure test to detect antibodies was compared to that of the STANDARD Q COVID-19 IgM/IgG Plus test and that of PRNT. The positive, negative, and overall percent agreement between the RapiSure and STANDARD Q test was 95.7%, 89.3%, and 91.5%, respectively, with a Cohen's kappa of 0.82. The RapiSure neutralizing antibody test results revealed a sensitivity of 93.4% and a specificity of 100% compared to the PRNT results, with an overall percent agreement of 97.5% and Cohen's kappa of 0.95. The diagnostic performance of the RapiSure test was in good agreement with the STANDARD Q COVID-19 IgM/IgG Plus test and comparable to that of the PRNT. The RapiSure S1 RBD IgG/Neutralizing Ab Test was found to be convenient and reliable and, thus, can provide valuable information for rapid clinical decisions during the COVID-19 pandemic.