Background: Magnetic resonance-guided focused ultrasound (MRgFUS) thalamotomy is used to treat essential tremor and tremor-dominant Parkinson disease. Feedback is collected throughout the procedure to verify the location of the target and completeness of response; however, variability in clinical judgments may underestimate or overestimate treatment response.
Objective: To objectively quantify joint motion after each sonication using accelerometers secured to the contralateral upper extremity in an effort to optimize MRgFUS treatment.
Methods: Before the procedure, 3 accelerometers were secured to the patient's arm, forearm, and index finger. Throughout the procedure, tremor motion was regularly recorded during postural and kinetic tremor testing and individual joint angle measures were modeled. The joint angle from each accelerometer was compared with baseline measurements to assess changes in angles. Subsequent adjustments to the target location and sonication energy were made at the discretion of the neurosurgeon and neuroradiologist.
Results: Intraoperative accelerometer measurements of hand tremor from 18 patients provided quantified data regarding joint angle reduction: 87.3%, 94.2%, and 86.7% for signature writing, spiral drawing, and line drawing tests, respectively. Target adjustment based on accelerometer monitoring of the angle at each joint added substantial value toward achieving optimal tremor reduction.
Conclusion: Real-time accelerometer recordings collected during MRgFUS thalamotomy offered objective quantification of changes in joint angle after each sonication, and these findings were consistent with clinical judgments of tremor response. These results suggest that this technique could be used for fine adjustment of the location of sonication energy and number of sonications to consistently achieve optimal tremor reduction.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1227/ons.0000000000000638 | DOI Listing |
Knee
January 2025
Department of Radiology, Keio University School of Medicine, Shinjuku, Tokyo, Japan.
Background: Long-leg alignment and joint line obliquity have traditionally been assessed using two-dimensional (2D) radiography, but the accuracy of this measurement has remained unclear. This study aimed to evaluate the accuracy of 2D measurements of lateral distal femoral angle (LDFA) and medial proximal tibial angle (MPTA) using upright three-dimensional (3D) computed tomography (CT).
Methods: This study involved 66 knees from 38 patients (34 women, four men) with knee osteoarthritis (OA), categorized by Kellgren-Lawrence (KL) grade.
BMC Musculoskelet Disord
January 2025
Department of Orthopedics, Peking University Third Hospital, No. 49. North Garden Street, Hai Dian District, Beijing, 100191, People's Republic of China.
Background: For degenerative lumbar scoliosis (DLS), prior studies mainly focused on the preoperative relationship between spinopelvic parameters and health-related quality of life (HRQoL), lacking an exhaustive evaluation of the postoperative situation. Therefore, the postoperative parameters most closely bonded with clinical outcomes has not yet been well-defined in DLS patients. The objective of this study was to comprehensively assess the correlation between radiographic parameters and HRQoL before and after surgery, and to identified the most valuable spinopelvic parameters for postoperative curative effect.
View Article and Find Full Text PDFSci Rep
January 2025
La Trobe Sport & Exercise Medicine Research Centre, School of Allied Health, Human Services and Sport, La Trobe University, Melbourne, VIC, Australia.
Few studies have explored hip morphology and cartilage composition in female athletes or the impact of asymmetric repetitive loading, such as occurs during softball pitching. The current cross-sectional study assessed bilateral bony hip morphology on computed tomography imaging in collegiate-level softball pitchers ('Pitch1', n = 25) and cross-country runners ('Run', n = 13). Magnetic resonance imaging was used to assess cartilage relaxation times in a second cohort of pitchers ('Pitch2', n = 10) and non-athletic controls ('Con', n = 4).
View Article and Find Full Text PDFPurpose: This study aimed to assess the posterior cruciate ligament (PCL) angle in anterior cruciate ligament (ACL) deficient knees and correlate it with anatomical and demographic factors such as tibial slope, anterior tibial translation, age, gender, and time of injury.
Material And Methods: Patients were eligible for inclusion if they were clinically diagnosed with an ACL tear confirmed by MRI. For each patient, the following parameters were evaluated: PCL angle (PCLA), medial tibial slope (MTS), lateral tibial slope (LTS), medial anterior tibial translation (MATT), and lateral anterior tibial translation (LATT).
J Biomech
January 2025
Department of Orthopedic Surgery and Orthopedic Research Institute, West China Hospital, Sichuan University Chengdu Sichuan Province China. Electronic address:
OpenCap, a smartphone-based markerless system, offers a cost-effective alternative to traditional marker-based systems for gait analysis. However, its kinematic measurement accuracy must be evaluated before widespread use in clinical practice. This study aimed to evaluate OpenCap for lower-limb joint angle measurements during walking in patients with knee osteoarthritis (OA) and to compare error metrics between patients and healthy controls.