A novel portable and disposable bipolar electrode (BPE)-electrochemiluminescence (ECL) device was fabricated for fumonisin B (FB) detection. BPE was fabricated by using MWCNTs and polydimethylsiloxane (PDMS) due to their excellent electrical conductivity and good mechanical stiffness. After the deposition of Au NPs on the cathode of BPE, the ECL signal could be improved 89-fold. Then a specific aptamer-based sensing strategy was constructed by grafting capture DNA on Au surface, followed by hybridizing with aptamer. Meanwhile, an excellent catalyst, Ag NPs was labeled on aptamer to activate oxygen reduction reaction, leading to a 13.8-fold enhancement in ECL signal at the anode of BPE. Under the optimal conditions, the biosensor exhibited a wide linear range of 0.10 pg/mL to 10 ng/mL for FB detection. Meanwhile, it demonstrated satisfactory recoveries for real sample detection with good selectivity, making it to be a convenient and sensitive device for mycotoxin assay.