Objectives: The diagnosis and monitoring of bleeding and thrombotic disorders depend on correct haemostatic measurements. The availability of high-quality biological variation (BV) data is important in this context. Many studies have reported BV data for these measurands, but results are varied. The present study aims to deliver global within-subject (CV) and between-subject (CV) BV estimates for haemostasis measurands by meta-analyses of eligible studies, by assessment with the Biological Variation Data Critical Appraisal Checklist (BIVAC).

Methods: Relevant BV studies were graded by the BIVAC. Weighted estimates for CV and CV were obtained via meta-analysis of the BV data derived from BIVAC-compliant studies (graded A-C; whereby A represents optimal study design) performed in healthy adults.

Results: In 26 studies BV data were reported for 35 haemostasis measurands. For 9 measurands, only one eligible publication was identified and meta-analysis could not be performed. 74% of the publications were graded as BIVAC C. The CV and CV varied extensively between the haemostasis measurands. The highest estimates were observed for PAI-1 antigen (CV 48.6%; CV 59.8%) and activity (CV 34.9%; CV 90.2%), while the lowest were observed for activated protein C resistance ratio (CV 1.5%; CV 4.5%).

Conclusions: This study provides updated BV estimates of CV and CV with 95% confidence intervals for a wide range of haemostasis measurands. These estimates can be used to form the basis for analytical performance specifications for haemostasis tests used in the diagnostic work-up required in bleeding- and thrombosis events and for risk assessment.

Download full-text PDF

Source
http://dx.doi.org/10.1515/cclm-2022-1207DOI Listing

Publication Analysis

Top Keywords

haemostasis measurands
16
biological variation
12
variation data
12
within-subject between-subject
8
studies graded
8
graded bivac
8
measurands
7
data
6
studies
5
estimates
5

Similar Publications

Objectives: The diagnosis and monitoring of bleeding and thrombotic disorders depend on correct haemostatic measurements. The availability of high-quality biological variation (BV) data is important in this context. Many studies have reported BV data for these measurands, but results are varied.

View Article and Find Full Text PDF

Objectives: A carrier prototype by Aerocom (Schwäbisch Gmünd, Germany) for pneumatic tube systems (PTS) is able to transport 9 blood tubes which are automatically fixed by closing the lid. In this study, we examined the influence of the transport on blood sample quality using the carrier prototype comparing to courier transport and a conventional carrier (AD160, Aerocom).

Methods: Triplicate blood samples sets (1 lithium heparin, 1 EDTA, 1 sodium citrate) of 35 probands were split among the transportation methods: 1.

View Article and Find Full Text PDF

Introduction: Physiologically, feline platelets are more reactive and prone to aggregation, which interferes with platelet counts using automated counters and manual methods. The use of aminoglycoside amikacin in association with EDTA has proven to be efficient in preventing platelet aggregates in cases of pseudo thrombocytopenia (PTP) in people.

Objectives: This study evaluated the efficacy of amikacin in preventing platelet aggregation in EDTA-containing feline blood samples and investigated the possible effects on hematologic measurands.

View Article and Find Full Text PDF

Top-down and bottom-up approaches for the estimation of measurement uncertainty in coagulation assays.

Clin Chem Lab Med

August 2020

Department of Laboratory Medicine, Chung-Ang University College of Medicine, 102 Heukseok-ro, Dongjak-Ku, Seoul 06973, Republic of Korea, Phone: +82-2-6299-2718, Fax: +82 2-6298-8630.

Background: The assessment of measurement uncertainty (MU) in clinical laboratories is essential to the reliable interpretation of results in clinical laboratories. However, despite the introduction of various methods for the expression of uncertainty in measurement, the MUs of coagulation tests have not been extensively studied. The aim of this study was to quantify the MU of various coagulation assays according to international guidelines and to report an expected confidence in the quality of coagulation assays.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!