Objective: Continuous glucose monitoring (CGM) may be challenged by extreme conditions during cardiac surgery using hypothermic extracorporeal circulation (ECC).

Research Design And Methods: We evaluated the Dexcom G6 sensor in 16 subjects undergoing cardiac surgery with hypothermic ECC, of whom 11 received deep hypothermic circulatory arrest (DHCA). Arterial blood glucose, quantified by the Accu-Chek Inform II meter, served as reference.

Results: Intrasurgery mean absolute relative difference (MARD) of 256 paired CGM/reference values was 23.8%. MARD was 29.1% during ECC (154 pairs) and 41.6% immediately after DHCA (10 pairs), with a negative bias (signed relative difference: -13.7%, -26.6%, and -41.6%). During surgery, 86.3% pairs were in Clarke error grid zones A or B and 41.0% of sensor readings fulfilled the International Organization for Standardization (ISO) 15197:2013 norm. Postsurgery, MARD was 15.0%.

Conclusions: Cardiac surgery using hypothermic ECC challenges the accuracy of the Dexcom G6 CGM although recovery appears to occur thereafter.

Download full-text PDF

Source
http://dx.doi.org/10.2337/dc22-2260DOI Listing

Publication Analysis

Top Keywords

cardiac surgery
16
surgery hypothermic
16
continuous glucose
8
glucose monitoring
8
hypothermic extracorporeal
8
extracorporeal circulation
8
hypothermic ecc
8
relative difference
8
surgery
5
hypothermic
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!